Dwyer Instruments GFM2 Series Installation And Operating Instructions Manual Download Page 10

7.3 Analog output Calibration of GFM2 Mass Flow Meters

GFM2  series  Mass  Flow  Meters  are  equipped  with
calibrated 0 to 5 Vdc and 4 to 20 mA output signals. The set
of the jumpers (J7A, J7B, J7C) on the printed circuit board
is used to switch between 0 to 5 Vdc and 4 to 20 mA output
signals (Figure C-1 “GFM2 configuration jumpers).

Figure C-1 GFM2 Analog Output Configuration

Jumpers

The  GFM2  analog  output  calibration  involves  calculation
and  storing  of  the  offset  and  span  variables  in  the
EEPROM for each available output. The 0 to 5 VDC output
has  only  scale  variable  and  20  mA  output  has  offset  and
scale  variables.  The  following  is  a  list  of  the  Gas
independent variables used for analog output computation:

Index 

Name 

Description

25 

AoutScaleV        - 

DAC 0 to 5 VDC Analog Output Scale

27 

AoutScale_mA    - 

DAC 4 to 20 mA Analog Output Scale

28 

AoutOffset_mA    - 

DAC 4 to 20 mA Analog Output Offset

7.3.1 Initial Setup

Power up the Mass Flow Meter for at least 15 minutes prior to commencing the
calibration procedure. Make sure absolutely no flow takes place through the
meter.  Establish  digital  RS-485/RS-232  communication  between  PC
(communication terminal) and GFM2. The commands provided below assume that
calibration  will  be  performed  manually  (w/o  Dwyer  supplied  calibration  and
maintenance software) and the device has RS-485 address 11. If Dwyer supplied
calibration and maintenance software is used, skip the next section and follow the
software prompts.

Enter Backdoor mode by typing:  !11,MW,1000,1[CR]
Unit will respond with: 

!11,BackDoorEnabled: Y

Disable DAC update by typing: 

!11,WRITE,4,D[CR]

Unit will respond with: 

!11,DisableUpdate: D

7.3.2 Gas Flow 0 to 5 VDC Analog Output Calibration

1. Install jumpers J7A, J7B and J7C on the PC board for 0-5 Vdc output (see
Table VI).
2. Connect a certified high sensitivity multi meter set for the voltage measurement
to the pins 2 (+) and 1 (-) of the 15 pins D connector.
3. Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]
4. Read voltage with the meter and calculate:

5. Save FlowOutScaleV in to the EEPROM: !11,MW,25,X[CR]

Where: X – the calculated AoutScaleV value.

7.3.3 Gas Flow 4 to 20 mA Analog Output Calibration

1. Install jumpers J7A, J7B and J7C on the PC board for 4 to 20 mA output (see
Table VI).
2. Connect a certified high sensitivity multimeter set for the current measurement
to pins 2 (+) and 1 (-) of the 15 pins D connector.
3. Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]
4. Read current with the meter and calculate:

5. Write zero counts to the DAC channel 1: !11,WRITE,1,0CR]
6. Read offset current with the meter and calculate:

7. Save AoutScale_mA in to the EEPROM: 

!11,MW,27,Y[CR]

Save AoutOffset_mA in to the EEPROM:  !11,MW,28,Z[CR]

Where:  Y – the calculated AoutScale_mA value

Z – the calculated AoutOffset_mA value

Enable DAC update by typing: 

!11,WRITE,4,N[CR]

Unit will respond with: 

!11,DisableUpdate: N

Close BackDoor access by typing:     !11,MW,1000,0[CR]
Unit will respond with:

!11,BackDoorEnabled: N

8. RS-485/RS-232 SOFTWARE INTERFACE COMMANDS
8.1 General

The standard GFM2 comes with an RS-485 interface. For the optional RS-232
interface, the start character (!) and two hexadecimal characters for the address
must be omitted. The protocol described below allows for communications with
the unit using either a custom software program or a “dumb terminal.” All values
are sent as printable ASCII characters. For RS-485 interface, the start character
is always (!). The command string is terminated with a carriage return (line feeds
are automatically stripped out by the GFM2). See section 2.2.3 for information
regarding communication parameters and cable connections.

8.2 Commands Structure

The structure of the command string:
!<Addr>,<Cmd>,Arg1,Arg2,Arg3,Arg4<CR>
Where:

Start character.

Addr 

RS485 device address in the ASCII representation of 
hexadecimal (00 through FF are valid).

Cmd 

The one or two character command from the table below.

Arg1 to Arg4 

The command arguments from the table below.
Multiple arguments are comma delimited.

CR Carriage 

Return character.

Several examples of commands follow. All assume that the GFM2 has been
configured for address 18 (12 hex) on the RS485 bus:
1. 

To get current calibration tables:

!12,G<CR> 

The GFM2 will reply: 

!12,G 0 AIR<CR>

(Assuming Current Gas table is #0, calibrated for AIR )

2. 

To get current Alarm status:

!12,A,R<CR>

The GFM2 will reply: 

!12,N<CR>

(Assuming no alarm conditions)

3. 

To get a flow reading: 

!12,F<CR>

The GFM2 will reply:

!12,50.0<CR> 

(Assuming the flow is at 50% FS)

4. 

Set the high alarm limit to 85% of FS flow rate:

!12,A,H,85.0<CR>

The GFM2 will reply:

!12,AH85.0<CR>

Analog
Output
RS485
Terminal Resistor

Function

0 to 5 VDC
4 to 20 mA

OFF

ON

J7A

5-9
1-5

J7B

6-10

2-6

J7C

7-11

3-7

JCD

8-12

4-8

The analog output available on the GFM2 Digital Mass Flow
Meter was calibrated at the factory for the specified gas and

full scale flow range (see the device’s front label). There is no need to perform
analog output calibration unless the EEPROM IC was replaced or offset/span
adjustment is needed. Any alteration of the analog output scaling variables in the
Gas independent table will VOID calibration warranty supplied with instrument.

NOTICE

It is recommended to use the Dwyer supplied calibration and
maintenance software for analog output calibration. This

software includes an automated calibration procedure which may radically
simplify calculation of the offsets and spans variables and, the reading and writing
for the EEPROM table.It is recommended to use the Dwyer supplied calibration
and maintenance software for analog output calibration. This software includes an
automated calibration procedure which may radically simplify calculation of the
offsets and spans variables and, the reading and writing for the EEPROM table.

NOTICE

AoutScaleV =

20000

Reading[V]

AoutScale_mA =

4000

Reading[mA]

AoutOffset_mA = -FlowOutScale_mA*Offset_Reading[mA]

When done with the analog output calibration make sure the
DAC update is enabled and the BackDoor is closed (see

command below).

NOTICE

Default address for all units is 11. Do not submit start
character and two character hexadecimal device address for

RS-232 option.

NOTICE

Page 9

Summary of Contents for GFM2 Series

Page 1: ...thin 2 of actual flow Output Signal Linear 0 to 5 VDC 3000 min load impedance and 4 to 20 mA 500 max loop resistance Relay Rating 1 A 24 VDC Max Particulate Size 5 microns Temperature Limits 32 to 122...

Page 2: ...ass Flow Meter Installation and Operating Instructions Bulletin F GFM2 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail...

Page 3: ...7 6 2 Flow Path Cleaning 7 6 2 1 Restrictor Flow Element RFE 7 6 2 2 GFM2 Models 7 7 Calibration Procedures 8 7 1 Flow Calibration 8 7 2 Gas Flow Calibration of GFM2 Mass Flow Meters 8 7 2 1 Connecti...

Page 4: ...thorough leak test of the entire system All GFM2 s are checked prior to shipment for leakage within stated limits See specifications in this manual 2 2 Electrical Connections GFM2 is supplied with a 2...

Page 5: ...nterface or optional Profibus DP interface GFM2 flow meters support various functions including programmable flow totalizer low high or range flow alarm automatic zero adjustment activated via local b...

Page 6: ...peration of the meter above 110 of maximum calibrated flow may increase recovery time 5 3 GFM2 Parameters Settings 5 3 1 Engineering Units Settings The GFM2 Mass Flow Meter is capable of displaying fl...

Page 7: ...nitrogen for up to 36 gases are stored in the GFM2 see APPENDIX II In addition provision is made for a user defined conversion factor Conversion factors may be applied to any of the ten gas calibrati...

Page 8: ...to disassemble the sensor If blockage of the sensor tube is not alleviated by flushing through with cleaning fluids please return meter for servicing 6 2 1 Restrictor Flow Element RFE The Restrictor...

Page 9: ...s the sensor Average reading must be in the range 120 10 counts If it is not perform Auto Zero procedure see section 5 3 10 Zero Calibration 7 2 3 Gas Linearization Table Adjustment Gas flow calibrati...

Page 10: ...RS 232 SOFTWARE INTERFACE COMMANDS 8 1 General The standard GFM2 comes with an RS 485 interface For the optional RS 232 interface the start character and two hexadecimal characters for the address mu...

Page 11: ...command can be used only when absolutely no flow thru the meter and no earlier then 6 minutes after power up It can take several minutes to complete Unit will not respond to other commands when this...

Page 12: ...R read current totalizer volume W warm up delay S setting status D disable sets K 1 I internal K factor U user specified factor S status FS mL sec mL min mL hr L sec L min L hr m3 sec m3 min m3 hr f3...

Page 13: ...w measurement LCD display reading does correspond to the correct flow range but 0 5 Vdc output signal does not change always the same reading or around zero LCD display reading and 0 to 5 Vdc output v...

Page 14: ...nsor temperature is too high MCU temperature is too high overload Fatal Error EEPROM or Auto Zero error Solution Flush clean or disassemble to remove impediments or replace the filter screen see secti...

Page 15: ...rent Gas Table Number 0 9 Current Units of Measure 0 22 Alarm Mode E Enabled D Disabled Low Flow Alarm Setting FS 0 Disabled High Flow Alarm Setting FS 0 Disabled Flow Alarm Action Delay 0 3600 sec 0...

Page 16: ...Flow in PFS 0 6 70 0 FS A D Value from Sensor counts Actual Flow in PFS 0 7 80 0 FS A D Value from Sensor counts Actual Flow in PFS 0 8 90 0 FS A D Value from Sensor counts Actual Flow in PFS 0 9 100...

Page 17: ...n 114 C2Cl2F4 Freon 115 C2ClF5 Freon C318 C4F8 Germane GeH4 Germanium Tetrachloride GeCl4 Helium He 50 L min Helium He 1 50 L min Helium He 2 10 50 L min Index 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16...

Page 18: ...butyl Aluminum C4H9 AL Titanium Tetrachloride TiCl4 Trichloro Ethylene C2HCl3 Trimethylamine CH3 3N Tungsten Hexafluoride WF6 Uranium Hexafluoride UF6 Vinyl Bromide CH2CHBr Vinyl Chloride CH2CHCl Xeno...

Page 19: ...APPENDIX IV Component Diagram Top Component Side Bottom Component Side Page 18...

Page 20: ...s required The Series GFM3 is not field serviceable and should be returned if repair is needed Field repair should not be attempted and may void warranty See 1 3 WARRANTY RETURN Refer to Terms and Con...

Page 21: ...wyer Instruments Inc Printed in U S A 5 13 FR RA 444052 00 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail info dwyer...

Reviews: