AUTO SERIAL CONFIGURATION 

Auto serial configuration enables the device to determine the baud rate, parity and 

stop bits directly from the serial traffic. This allows a device to be quickly and easily 

deployed after a valid RS-485 MAC address is chosen. Note that the auto configuration 

procedure  assumes  a  serial  configuration  appropriate  to  the  selected  protocol  as 

follows:

When Modbus

®

 communications protocol is selected, and the device is setup offline or 

away from the main network, it is necessary to generate traffic in order to configure the 

serial communication. Attempting to read input registers is a good method to generate 

traffic. Note that while serial configuration is in progress, the device may not respond 

to requests. The device may require multiple read requests to complete the serial 

configuration process. 

The auto serial configuration process will complete once a message addressed to the 

device is received and processed successfully. The serial configuration parameters 

are then saved to non-volatile storage and loaded by default each time the device 

starts. If the serial configuration of the bus changes, a power cycle of the device is 

required to restart the auto serial configuration process.

BACnet Communication Protocol Services

Device Communication Control Service (DM-DCC-B) 

This device supports the Device Communication Control Service BIBB. The optional 

time duration in minutes is also supported. This device is configured with a password 

that must be provided to successfully execute this command.

  The  password  is 

“Dwyer”. 

Reinitialize Device Service (DM-RD-B) 

This device supports the Reinitialize Device Service BIBB. The supported device 

states are COLDSTART and WARMSTART. All other states return error. This device 

is  configured  with  a  password  that  must  be  provided  to  successfully  execute  this 

command. 

The password is “Dwyer”.

BACnet Communication Protocol Objects Overview

The instrument supports the following objects:

SUPPORTED BACNET COMMUNICATION PROTOCOL OBJECTS

Object Type

Dynamically 

Creatable

Dynamically 

Deletable

Object 

Identifier Object Name

Device

Analog Input

Analog Value

Analog Value 

Analog Value 

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Analog Value

Multi-State Value

Multi-State Value

Multi-State Value

Multi-State Value

Multi-State Value

Multi-State Value

Multi-State Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Binary Value

Positive 

Integer Value

Positive 

Integer Value

Positive 

Integer Value

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

607xxx

AI1

AV1

AV2

AV3

AV4

AV5

AV6

AV7

AV8

AV9

AV10

AV11

AV12

AV13

AV14

MSV1

MSV2

MSV3

MSV4

MSV5

MSV6

MSV7

BV1

BV2

BV3

BV4

BV5

BV6

BV7

BV8

BV9

BV10

BV11

BV12

PIV1

PIV2

PIV3

DHC 607xxx

Pressure

Velocity

Flow

Peak

Valley

Velocity K-Factor

Flow Area

Process Output Low

Process Output High

Set Point 1 Trigger Value

Set Point 1 Reset Value

Set Point 2 Trigger Value

Set Point 2 Reset

Alarm High limit

Alarm Low Limit

Process Mode

Voltage Output Range

Relay 1 Source

Relay 1 Action

Relay 2 Source

Relay 2 Action

External Input Type

Enable Set Point 1

Set Point 1 Status

Enable Set Point 2

Set Point 2 Status

Enable Alarm High Limit

Alarm High Limit Status

Enable Alarm Low Limit

Alarm Low Limit Status

Manual Alarm Reset

Power Up Alarm Inhibit

Enable External Input

External Input Status

Alarm Delay

External Input Activation 

Delay

Damping Time

Note: 

The default object identifier is 607xxx, where xxx is replaced by the MS/TP 

MAC address set in the ADDR SUB MENU under the Conn MENU. The object 

identifier  value  will  change  as  the  MS/TP  MAC  address  changes.  However,  if  a 

specific  object  identifier  is  written  via  BACnet  communication  protocol,  then  that 

value is stored and changes to the MS/TP MAC address will no longer affect the 

object identifier. Similarly, the default object name includes 607xxx. The object name 

will reflect the current object identifier. If a specific object name is written via BACnet 

communication protocol, then that value is stored and changes to the object identifier 

will no longer affect the object name.

SUPPORTED SERIAL CONFIGURATIONS

Protocol

Supported 

Baud Rates

Data Size

Parity

Stop Bits

BACnet - MS/TP

9600

19200

38400

57600

78600

115200

8

None

1

Modbus

®

 - RTU

Even

Odd

None

1

None

2

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Summary of Contents for Digihelic DHC Series

Page 1: ...this device in a dusty environment install an optional A 331 filter vent plug into the vented port to keep interior clean Differential Pressure Connect tubing from the higher source to the HIGH PRESSURE port indicated with a sign and from the lower source to the LOW PRESSURE port indicated with a sign SPECIFICATIONS Service Air and non combustible compatible gases Wetted Materials Consult factory ...

Page 2: ...urface Provide three 3 16 diameter holes in panel on a 4 1 8 diameter bolt circle Cut the opening for the terminal block as shown in Figures 2 and 4 2 If the rear pressure connections are to be used also provide 1 2 diameter holes as shown in Figures 2 and 4 3 Insert 6 32 machine screws from rear of mounting surface threading them into tapped holes on back of transmitter and tighten 3 HOLES FOR 6 ...

Page 3: ...C 1 1 2 38 10 3 4 19 05 7 8 22 23 1 25 40 53 64 20 96 1 25 40 2X Ø5 8 15 88 FOR PRESSURE CONNECTION DRILL TEMPLATE NEEDS TO BE USED FROM PANEL FRONT Figure 4 Surface mounting drill template www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 4: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 5: ...ire in accordance with an equivalent national standard or code Use copper conductors rated for 60 C All terminals are rated CLASS 2 CAUTION Function 15 Pin Connector Terminal Cable Color 12 28 VAC VDC Power 12 28 VAC VDC Power 4 20 mA Output Signal 0 10 V Output Signal External Input Signal Signal Common Relay 1 N O Relay 1 Common Relay 1 N C Relay 2 N O Relay 2 Common Relay 2 N C RS 485 B RS 485 ...

Page 6: ...voltage output Set POL to the desired display reading for 4 mA output Set POH to the desired display reading for 20 mA output POL can be greater than POH In this situation the analog outputs are reverse acting KFAC SUB MENU The K Factor sub menu becomes accessible if the instrument mode is set to Velocity or Flow When the DHC is used with a Pitot tube the manufacturer may specify a K Factor The ad...

Page 7: ...2 Action rLy1 rLy2 Relay 1 Source Relay 2 Source SUB MENUS OFF Relay not used SP1 Set Point 1 SP2 Set Point 2 AL Either Alarm ALHI High Alarm ALLO Low Alarm EXT External Input The selected source determines when the relay is active If the source is active the relay will be active r1At r2At Relay 1 Action Relay 2 Action SUB MENUS DIR Direct Acting CLSE Close on Active REV Reverse Acting OPEN Open o...

Page 8: ...ue Press and hold the ENTER button until the process value is cleared The display will show a dash in each digit starting with the left most position After 4 dashes have been displayed the process value will be displayed and the new span will be applied If the span cannot be applied a FAiL message will be displayed This error could occur if the applied pressure is too low or too high rSt Reset Fac...

Page 9: ...log Value Analog Value Analog Value Analog Value Analog Value Multi State Value Multi State Value Multi State Value Multi State Value Multi State Value Multi State Value Multi State Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Binary Value Positive Integer Value Positive Integer Value Positive I...

Page 10: ...BACnet Services Supported BACnet Object Types Supported BACnet Array Unsigned BACnet Segmentation Unsigned Unsigned Unsigned Unsigned BACnet Address Binding Unsigned BACnet Array N of BACnet Property Identifier Character String Read Write Read Write Read Read Read Read Read Read Read Read Write Read Write Read Read Read Read Read Read Read Read Read Read Write Read Read Read Read Read Property Def...

Page 11: ...tus Flags BACnet Event State BACnet Reliability Boolean BACnet Engineering Units Real Real Read Read Read Read Write Read Read Read Read Write Read Read Read Property Default Value Property Data Type Access Object Identifier Object Name Object Type Preset Value Status Flags Event State Reliability Out Of Service Units Max Pres Value Min Pres Value AV6 Flow Area Analog Value 1 0 0 093 0 0 0 0 Norma...

Page 12: ...tatus Flags Event State Reliability Out Of Service Units Max Pres Value Min Pres Value AV9 Set Point 1 Trigger Value Analog Value Model Specific 0 0 0 0 Normal No Fault Detected FALSE Mode Specific Model Specific Model Specific BACnet Object Identifier Character String BACnet Object Type Real BACnet Status Flags BACnet Event State BACnet Reliability Boolean BACnet Engineering Units Real Real Read ...

Page 13: ...ject Identifier Object Name Object Type Preset Value Status Flags Event State Reliability Out Of Service BV5 Enable Alarm High Limit Binary Value Active 0 0 0 0 Normal No Fault Detected FALSE BACnet Object Identifier Character String BACnet Object Type BACnet Binary PV BACnet Status Flags BACnet Event State BACnet Reliability Boolean Read Read Read Read Write Read Read Read Read Write Property Def...

Page 14: ...perty Data Type Access Object Identifier Object Name Object Type Preset Value Status Flags Event State Reliability Out Of Service Number of States State Text MSV3 Relay 1 Source Multi State Value 2 0 0 0 0 Normal No Fault Detected FALSE 7 Off Set Point 1 Set Point 2 Any Alarm Alarm High LImit Alarm Low Limit External Input BACnet Object Identifier Character String BACnet Object Type Unsigned BACne...

Page 15: ...ad Read Read Write Read Read Read Read Write Property Default Value Property Data Type Access Object Identifier Object Name Object Type Preset Value Status Flags Event State Reliability Out Of Service Units Max Pres Value Min Pres Value PIV2 External Input Activation Delay Positive Integer Value 0 0 0 0 0 Normal No Fault Detected FALSE Mode Specific 3600 0 BACnet Object Identifier Character String...

Page 16: ...able Set Point 2 Trigger Value Set Point 2 Reset Value Enable Alarm High Limit High Alarm Limit Enable Alarm Low Limit Low Alarm Limit Manual Alarm Reset Power Up Alarm Inhibit Alarm Delay Alarm High Limit Status Alarm Low Limit Status Relay 1 Source Relay 1 Action Relay 2 Source Relay 2 Action Enable External Input External Input Type External Input Activation Delay Damping Time String 32 16bit i...

Page 17: ...pted and may void warranty WARRANTY RETURN Refer to Terms and Conditions of Sale in our catalog and on our website Contact customer service to receive a Return Materials Authorization RMA number before shipping the product back for repair Be sure to include a brief description of the problem plus any additional application notes Byte Order Address Range 32bit Values 16bit Values Register 1 Registe...

Page 18: ...s MODEL PRESSURE RANGE 3s HOME SCREEN MODE PRESSURE PRESS AND HOLD ENTER BUTTON PRESS SPAL BUTTON PRESS UP BUTTON PRESS DOWN BUTTON PRESS RST BUTTON PIN CORRECT NO NEXT DIGIT MOVE CARON OR CHANGE DIGIT OR PIN ENABLED YES MAIN MENU SPAL MENU OR 3s MENU MAPS www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 19: ...ARM MENU RELAY MENU EXTERNAL INPUT MENU DISPLAY MENU CONNECTION MENU CALIBRATION MENU HOME SCREEN NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD 0000 DISABLED CONFIGURE PIN www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 20: ...OPERATION MENU MAIN MENU OPERATION MODE MEASUREMENT UNIT OPERATION MENU CONTINUED www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 21: ...VE CARON OR CHANGE DIGIT SAVE DISCARD OPERATION MENU CONTINUED FLOW OUTPUT HIGH FLOW OUTPUT LOW DUCT AREA VELOCITY OUTPUT HIGH VELOCITY OUTPUT LOW VELOCITY K FACTOR PRESSURE OUTPUT HIGH PRESSURE OUTPUT LOW AVAILABLE WHEN MODE PRESSURE AVAILABLE WHEN MODE VELOCITY AVAILABLE WHEN MODE FLOW AVAILABLE WHEN MODE VELOCITY OR FLOW www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 22: ...OPERATION MENU CONTINUED VOLTAGE OUTPUT RANGE www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 23: ...SET POINT 1 ENABLE SET POINT 2 ENABLE MAIN MENU SET POINT MENU www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 24: ...ENABLE ALARM LOW LIMIT ENABLE MAIN MENU ALARM RESET TYPE ALARM INHIBIT ENABLE ALARM DELAY NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD ALARM MENU www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 25: ...MAIN MENU RELAY MENU RELAY 2 SOURCE RELAY 1 SOURCE RELAY MENU CONTINUED www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 26: ... OR EXT RELAY MENU CONTINUED RELAY 1 ACTION IF RELAY 1 SOURCE AL ALHI OR ALLO RELAY 2 ACTION IF RELAY 2 SOURCE SP1 SP2 OR EXT RELAY 2 ACTION IF RELAY 2 SOURCE AL ALHI OR ALLO www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 27: ...NAL INPUT MENU EXTERNAL INPUT ENABLE EXTERNAL INPUT TYPE MAIN MENU EXTERNAL INPUT DELAY NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 28: ...SS VALUE PROCESS DISPLAY MAIN MENU DAMPING TIME NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD VALLEY PROCESS VALUE RESET PEAK VALUE RESET VALLEY VALUE www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 29: ...ELECTION DEVICE ADDRESS MAIN MENU NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD CONNECTION MENU SERIAL BAUD RATE CONNECTION MENU CONTINUED www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 30: ...OPERATION MENU CONTINUED SERIAL PARITY AVAILABLE WHEN PROTOCOL SELECTION MODBUS AND SERIAL BAUD RATE AUTO SERIAL STOP BITS www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 31: ...ENU 1sec 1sec 1sec 1sec SPAN CALIBRATION 1sec 1sec 1sec 1sec DEVICE REBOOTS AFTER 5 SECONDS RESTORE FACTORY SETTINGS SERIAL NUMBER DISPLAY PART 1 SERIAL NUMBER DISPLAY PART 2 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Page 32: ...ARD NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD NEXT DIGIT MOVE CARON OR CHANGE DIGIT SAVE DISCARD ALARM LOW LIMIT ALARM HIGH LIMIT AVAILABLE WHEN ALARM LOW LIMIT ENABLE ON AVAILABLE WHEN ALARM HIGH LIMIT ENABLE ON Printed in U S A 6 22 FR 444618 00 Copyright 2022 Dwyer Instruments Inc DWYER INSTRUMENTS INC P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Phone 219 879 8000 Fax 219 872 9057 ww...

Reviews: