17
ALrE
(Alarm Reset) SUB MENU
ONOF
- Automatic reset
HOLD
- Manual reset. An alarm is reset by the RESET key on the front panel, or an external
reset switch.
AL.H
(Low Alarm Inhibit) SUB MENU
ON
- Alarm inhibit is on
OFF
- Alarm inhibit is off
If
AL.H
is selected ON, a low alarm condition is suspended upon power up until the process
value passes through the alarm set point once.
ALDL
(Alarm Delay) SUB MENU
Sets the amount of time an alarm condition must be continuously met before the alarm
condition is recognized. The alarm delay is adjustable from 0-3600 seconds.
d.S
(Display) MAIN MENU
PEAK
- Peak value
rESO
- Resolution
VALy
- Valley value
Pd.S
- Process display
ZERO
- Zero
DAMP
- Dampening level
PEAK
(Peak) SUB MENU
The Peak feature stores the highest pressure reading the instrument has measured since the
last reset or power up. At power up
PEAK
is reset to the present pressure reading. To manually
reset the
PEAK
value, press the RESET key while in the
PEAK
SUB MENU.
VALy
(Valley) SUB MENU
The valley feature stores the lowest pressure reading the instrument has measured since the
last reset or power up. At power up
VALy
is reset to the present pressure reading. To
manually reset the
VALy
value, press the RESET key while in the
VALy
SUB MENU.