10
1015100100L02 1910V003
Product description
10
9
24
12
11
8
Figure 2: SICOLAB mini ENDO – view of the
front
8 Quick-release coupling
9 On/Off switch
10 Pressure gauge
11 Unit cover
12 Unit cover lock
24 Fine filter 0.01 µm
13
14
15
Figure 3: View of the rear
13 Carry handle
14 Condensate drain
15 Power plug with fuses
5 Operation
The unit draws in atmospheric air and com-
presses it. The oil-free compressed air is then
transported to the pressure vessel. The oil-free
and filtered air is made available to the consum-
ers in the pressure vessel.
If compressed air is removed for a consumer,
the pressure in the vessel drops. When the
switch-on pressure is reached, the unit is auto-
matically switched on again via the pressure
switch. When the cut-off pressure is reached,
the unit is automatically switched off. A safety
valve prevents the maximum permissible vessel
pressure from being exceeded.
5.1 Unit details
1
2
3
4
6
5
7
Figure 1: Top view (without unit cover)
1 Pressure vessel
2 Safety Valve
3 Pressure switch
4 Air intake filter
5 Pressure reducer with condensate separa-
tor and fine filter 5 µm
6 Soleinoid valve
7 Operating time counter
EN