![Durr Dental XR 24 Installation And Operating Instructions Manual Download Page 14](http://html.mh-extra.com/html/durr-dental/xr-24/xr-24_installation-and-operating-instructions-manual_2542952014.webp)
14
14
11.3 XR 24 PRO
The regeneration function ON / OFF and the
regeneration time can be read on the display
of the X-ray developer (Fig. 14).
Please refer to the Installation and Operating
Instructions of the X-ray developer
XR 24 PRO (order no. 9000-600-74/30)
See section 18.6 Service Routine.
•
•
•
•
•
Start Service Routine
Switch off appliance at main switch
(press min. 2 s)
Simultaneously press
+
+
+
+
+
(min. 4s) Appliance is in service set-up
routine
•
•
•
•
•
Service function selection
Using keys and display select
REGENERATE, then select REGEN.ON.
•
Alter setting
(Factory setting 30s)
or
changes the parameter value
confirms new value
•
•
•
•
•
Finish service set-up routine
Switch off appliance at main switch
(press min. 2 s)