Assembly
Add device
Requirements:
– Unit switched on and connected to the
network
– Administrator or service technician access
level selected in the software
i
Working in the menu bar, click on
Device
Manager
.
The list of units appears. A symbol displays
the connection status to the software:
The device is present in the network
and connected to the software.
The device is present in the network
but not connected to the software.
The network connection between the
software and the device is interrupt-
ed, e.g. the device is switched off.
The new unit that is not yet connected, is
displayed with the connection status .
i
Select the unit and click on
.
The unit appears in the side bar.
Network configuration
Various options are available for network config-
uration:
– Automatic configuration via DHCP (recom-
mended).
– Automatic configuration via Auto-IP for direct
connection of unit and computer.
– Manual configuration.
i
Configure the network settings of the unit us-
ing the software or, if available, the touch
screen.
i
Check the firewall and release the ports, if ap-
plicable.
Further information on Tyscor Pulse can
be found in the software help and in the
Tyscor Pulse manual, order number
0949100001.
Network protocols and ports
Port
Purpose
Serv-
ice
45123 UDP,
45124 UDP
Unit recognition and
configuration
1900 UDP
Service indicator
SSDP /
UPnP
502 TCP
Unit data
514
1)
UDP
Event protocol data
Syslog
22 TCP,
23 TCP
Diagnosis
Telnet,
SSH
123 UDP
Time
NTP
1)
The port can vary depending on the config-
uration.
The following requirements must be met in order
to monitor the unit with the software on the
computer:
– Unit connected to the network
– Latest Tyscor Pulse software installed on
computer
EN
Summary of Contents for Power Tower Silence 200
Page 1: ...EN 1806V002 A949100002L02 Installation and operating instructions Power Tower Silence 200 ...
Page 2: ......
Page 25: ...A949100002L02 1806V002 23 11 Connection media plan PTS 200 Appendix EN ...
Page 26: ...24 A949100002L02 1806V002 Appendix PTS 200 PC board mit with CA 4 EN ...
Page 27: ...A949100002L02 1806V002 25 Appendix EN ...
Page 33: ...A949100002L02 1806V002 31 Appendix Y1 Relief valve EN ...
Page 34: ...32 A949100002L02 1806V002 Appendix X3 Top display optional X4 Manifold control contact EN ...
Page 38: ...36 A949100002L02 1806V002 Appendix M5 Drying unit fan Y1 Relief valve Q4 Pressure switch EN ...
Page 39: ...A949100002L02 1806V002 37 Appendix X4 24 V AC Ablagesteuerkontakt EN ...
Page 42: ...40 A949100002L02 1806V002 Appendix Y1 Relief valve EN ...
Page 43: ...A949100002L02 1806V002 41 Appendix X4 24 V AC manifold control contact EN ...
Page 46: ...44 A949100002L02 1806V002 Appendix EN ...
Page 47: ......