18 Tips for operators and service technicians
Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our
service.
Prior to working on the unit or in case of danger, disconnect it from the mains.
18.1 Displaying the sealing parameters
❯
Tap
Sealing Parameters > Live Display
.
❯
Feed through a test foil.
In this mode the device settings, such as the sealing temperature, sealing speed and contact pres-
sure, are displayed in real time.
18.2
Fault on the unit
Error
Possible cause
Remedy
Clicking noise during sealing
process.
Transport roller dirty.
❯
Inform your Service Techni-
cian.
18.3 Error messages on the touch screen
Error
Possible cause
Remedy
No SD card present.
Insert the SD card and restart
the device.
No SD card has been inserted.
❯
Insert an SD card when the
unit is switched off.
❯
If no SD card is to be used,
deactivate the function in the
menu.
SD card not detected.
❯
Remove the SD card with the
unit switched off and re-insert
it.
SD card write-protected.
❯
Remove the write protection.
SD card defective.
❯
Replace the SD card with a
new one.
SD card write error.
Restart the device.
SD card defective.
❯
With the device switched off,
replace the SD card with a
new one.
❯
Restart the device.
SD card full.
Insert an empty SD card and
restart the device.
Maximum data volume reached.
❯
Save data on PC.
❯
Insert an empty SD card and
restart the device.
Device overheating.
Please wait.
Inadequate ventilation of the
unit.
❯
Acknowledge the message.
Provide sufficient ventilation.
Wait until the temperature has
settled within the required
range. In the event of a per-
sistent fault, contact a Service
Technician.
Defective heating system.
Troubleshooting