14
Troubleshooting
Problem
Possible Reason
Solution
The SpO
2
and Pulse
Rate is not displaying
normally
a.The finger is not
properly positioned.
b.The patient’s SpO
2
is
too low to be detected.
a.Place the finger
properly and try again.
b.Try again; If you are
sure that the device
is working well, seek
medical help for a
diagnosis.
The SpO
2
and Pulse
Rate are not displayed
stably
a.The finger is not
placed inside deep
enough.
b.The finger is shaking
or the patient is moving.
a.Place the finger
properly and try again.
b.Calm the patient and
then try again.
The device cannot be
turned on
a.Low battery or no
battery.
b.The batteries are not
inserted properly.
c.The malfunction of the
device.
a.Change batteries.
b.Check batteries are
installed correctly.
c.Contact Duronic.
The display has turned
off suddenly
a.The device will power
off automatically when
there is no signal within
16 seconds.
b.The batteries are
almost drained.
a.Normal.
b.Change batteries.
Authorized European Representative:
Wellkang Ltd
Manufactured by:
Shenzhen Jumper Medical
Equipment Co. Ltd
Distributed by:
SHINEMART Ltd.
Summary of Contents for OX01R
Page 1: ...OX01R Pulse Oximeter PLEASE READ ALL THE INSTRUCTIONS BEFORE USE ...
Page 2: ...2 ...
Page 4: ...4 ...
Page 5: ...5 English Manual ...
Page 18: ...18 Manuel d utilisation en français ...
Page 20: ...20 Attention ...
Page 30: ...30 Gebrauchsanleitung auf Deutsch ...
Page 42: ...42 Manual De Instrucciones En Español ...
Page 54: ...54 Manuale D istruzioni In Italiano ...
Page 66: ...66 Instrukcja obsługi Polski ...
Page 78: ...78 Gebruikershandleiding Nederlands ...
Page 90: ......