6

DIODE MEASUREMENT

• 

Move the rotary dial to the       position and switch to the diode function.

• 

Contact the red probe to the anode of the measured diode, and the black probe to 

the cathode of the measured diode.

• 

The reading on the screen is the approximate value of the diode forward voltage 

drop. If it is connected in reverse, it will display “OL”.

       

Warning

• 

Firstly, disconnect the power supply of the circuit and discharge all high voltage 

capacitors.

• 

If the diode to be tested is an open circuit or with reverse polarity, the instrument 

will display “OL”.

• 

After completing all measurements, always disconnect the probe and the 

measured circuit.

 

CONNECTIVITY TEST

• 

Move the rotary dial to the       position, press “FUNC.” and switch to the 

connectivity function.

• 

Contact the probe to both ends of the measured circuit. When the resistance of the 

measured circuit is below 50Ω, the buzzer will sound.

       

Warning

• 

Firstly, disconnect the power supply of the circuit to be measured and fully 

discharge all high voltage capacitors.

• 

After completing all measurements, always disconnect the probe and the 

measured circuit.

 

NON-CONTACT VOLTAGE DETECTION

• 

Move the rotary knob of the instrument to NCV.

• 

Approach the non-contact voltage induction area of the instrument to the live wire 

of the AC voltage (less than 5mm).

• 

The non-contact voltage indicating light of the instrument will light up. At the same 

time there is a ‘beep’, which indicates that there is AC voltage on the fire wire.

GENERAL TECHNICAL SPECIFICATION

Environmental conditions 

for use

IEC/EN 61010-1 600V CAT III, pollution level 2, altitude 

<2000m

Working environment 

temperature & humidity

0~40°C

Storage environment 

temperature & humidity

-10~60°C (remove the battery when it is below 70%RH)

Temperature coefficient

0.1 x accuracy/°C

Max. Voltage allowed 

between measuring 

terminal and ground

DC or AC 600V RMS

Fuse Protection

Fuse FF 250mA/600V

Sampling rate

About 3 times/second

Summary of Contents for D03127

Page 1: ...1 MODEL D03127 Digital Multimeter Pen Type...

Page 2: ...rview 4 Measurement Operation 5 AC DC Voltage Measurement 5 AC DC Current Measurement 5 Resistance Measurement 6 Diode Measurement 6 Connectivity Test 6 Non contact Voltage Detection 6 General Technic...

Page 3: ...instrument around explosive gas steam or in a wet environment When measuring connect the zero ground line first then connect the live wire When disconnecting ensure you disconnect the live wire first...

Page 4: ...unit is turned the default mode is automatic range on or the function is switched the instrument displays AUTO symbol The operations of entering or quitting manual range are as follows In automatic r...

Page 5: ...screen When measuring DC current the display also shows the voltage polarity of the red probe test point Note If OL is displayed it means that the instrument exceeds the measurement range of the instr...

Page 6: ...50 the buzzer will sound Warning Firstly disconnect the power supply of the circuit to be measured and fully discharge all high voltage capacitors After completing all measurements always disconnect t...

Page 7: ...humidity is no more than 80 DC Voltage Range Resolution Accuracy 200mV 0 1mV 0 5 reading 3 2V 0 001V 20V 0 01V 200V 0 1V 600V 1V 0 8 reading 5 Input impedance 10M Maximum input voltage DC or AC 600V...

Page 8: ...ore than about 50 Testing current about 1mA open circuit voltage about 2 8V Input protection Maximum DC or AC 600V RMS REPLACING THE BATTERIES Turn off the power supply and disconnect the probe from t...

Reviews: