D U N I W A Y S T O C K R O O M C O R P .
20 of 29
Displayed pressure
Analog Output - volts
LO
0.00
0 mTorr (µBar) or less
0.00
0.016 mTorr (µBar)
0.10
0.10 mTorr (µBar)
0.50
0.20 mTorr (µBar)
0.65
1.0 mTorr (µBar)
1.00
1.58 mTorr (µBar)
1.10
10.0 mTorr (µBar)
1.50
100 mTorr (µBar)
2.00
1.00 torr (mBar)
2.50
10.0 Torr (µBar)
3.00
100 Torr (mBar)
3.50
999 Torr (mBar)
4.00
10.00 kTorr (mBar)
4.50
OFF
or
HI
5.00
Figure 13: Analog Output and Displayed Pressure
The pressure as a function of the Analog Output voltage is:
P=0.01*log
-1
(2V)
or
P=0.01*10
(2V)
NOTE:
The analog output is valid for the gauge which is selected on the display.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: [email protected]