Du Pont SORVALL RT6000B Operating Instructions Manual Download Page 3

Important Safety Information

Certain potentially dangerous conditions are inherent to the use of all centrifuges.  To
ensure safe operation of this centrifuge, anyone using it should be aware of all safe
practices and take all precautions described below and throughout this manual.

WARNING

When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of the
materials and the hazards associated with them in the event leakage occurs during
centrifugation.  If leakage does occur, neither the centrifuge nor the rotor can protect you
from particles dispersed in the air.  To protect yourself, we recommend additional precautions
be taken to prevent exposure to these materials, e.g., use of controlled ventilation or isolation
areas.

Always be aware of the possibility of contamination when using radioactive, toxic, or
pathogenic materials.  Take all necessary precautions and use appropriate decontamination
procedures if exposure occurs.

The use of sealed rotors, buckets and/or sample containers will provide increased protection
from contamination during routine operation.  However, these items will not guarantee
contamination protection from accidents resulting in damage to the rotor or buckets.  Do not
run hazardous materials in the centrifuge unless placed in a biohazard enclosure and operated
using all appropriate safety precautions.

Never use any material capable of producing flammable or explosive vapors.

Never exceed the maximum rated speed of the installed rotor; to do so can cause rotor failure.

Always reduce (derate) rotor speed as instructed in this manual whenever:

the rotor speed/temperature combination exceeds the solubility of the gradient material
and causes it to precipitate.

the compartment load exceeds the maximum allowable compartment load specified.  See
Chapter 3, Operation, page 3-6.

Failure to reduce rotor speed under these conditions can cause rotor failure.

CAUTION

Do not run or precool a rotor at the critical speed, as this will have a detrimental effect on
centrifuge component life.  See Chapter 3, page 3-3.

Do not operate the centrifuge with a rotor out of balance.  To do so can cause damage to the
centrifuge drive assembly.

Do not operate centrifuge without a rotor properly installed: rotor cover must be on and
locked in place, and the rotor locked to the centrifuge drive.  See rotor instruction manual.

The centrifuge can be damaged if connected to the wrong voltage.  Check the voltage before
plugging the centrifuge into a power source.  Du Pont is not responsible for incorrect
installation.

Always maintain the centrifuge in the recommended manner.  See Chapter 4, Maintenance.

Safety

SORVALL

®

 

Centrifuges

iii

Summary of Contents for SORVALL RT6000B

Page 1: ...SORVALL Centrifuges PN 07903 3 Issued July 1993 Du Pont Company SORVALL Products Wilmington Delaware 19898 U S A OPERATING INSTRUCTIONS SORVALL RT6000B T6000B TABLETOP CENTRIFUGES...

Page 2: ...icense to operate under nor a recommen dation to infringe upon any process patents Publications prior to the Issue Date of this manual may contain data in apparent conflict with that provided herein P...

Page 3: ...zardous materials in the centrifuge unless placed in a biohazard enclosure and operated using all appropriate safety precautions Never use any material capable of producing flammable or explosive vapo...

Page 4: ...n the RT6000B Centrifuge 3 3 Running Hazardous Material 3 4 Operation 3 4 Emergency Sample Recovery 3 6 Reducing Speed for Rotor Compartment Loads in Excess of Design Mass 3 6 Chapter 4 MAINTENANCE In...

Page 5: ...1 3 2 Location of Mechanical Override Loop 3 6 4 1 REV MIN x 1000 Indicator Adjustment 4 3 4 2 Motor Brush Assembly 4 4 4 3 CHAMBER TEMP C Indicator Adjustment 4 4 List of Tables Table Page 3 1 Contro...

Page 6: ...TOGRAPH OF THIS INSTRUMENT WAS NOT TAKEN AND THEREFORE WILL NOT BE AVAILABLE FOR FIRST PRINTING OF MANUAL Figure 1 1 SORVALL RT6000B Refrigerated Tabletop Centrifuge T6000B not shown SORVALL Centrifug...

Page 7: ...ted tabletop centrifuge Both the RT6000B and T6000B feature a closed loop speed control and have a see through chamber door that allows rotor calibration and visual inspection of a run in progress The...

Page 8: ...ors available for the RT6000B and T6000B Centrifuges For more information about other rotor accessories and tubes refer to the most current SORVALL Rotors Tubes Bottles and Adapters Catalog Rotor Spec...

Page 9: ...B 115 Vac 60 Hz 8A single phase 220 Vac 50 Hz 4A single phase Dimensions Width RT6000B 81 cm 32 inches T6000B 66 cm 26 inches Depth 64 cm 25 inches Height 32 cm 12 5 inches Mass Weight RT6000B 85 kg 1...

Page 10: ...lifting techniques can result in possible damage to the centrifuge and or personal injury CAUTION Do not lift the centrifuge by the front panel or by the lid To do so can result in damage to these pa...

Page 11: ...e plug will fit Hubbell receptacle No 5261 for 115 volt parallel blades or receptacle No 5561 for 220 volt tandem blades The power cord must be attached when the centrifuge is installed Installation T...

Page 12: ...controls and indicators Table 3 1 lists them by item number and describes their function Figure 3 1 Controls and Indicators Table 3 1 Controls and Indicators keyed to figure 3 1 Item Name Function 1...

Page 13: ...un time up to 30 minutes or 35 minutes on 50 Hz instruments At sets indefinite run time The TIME dial acts as a start switch to begin centrifuge run 8 Door release knob When the knob is turned to OPEN...

Page 14: ...he rotor be sure not to exceed the maximum compartment mass of the rotor see Table 1 1 or the individual rotor instruction manual If maximum compartment mass is exceeded maximum rotor speed must be lo...

Page 15: ...released into the laboratory area For this reason when materials which are pathogenic toxic or otherwise hazardous in nature are to be run the centrifuge should be placed in a biohazard safety enclos...

Page 16: ...peed NOTE The settings on the REV MIN x 1000 dial indicate actual rotor speed Verify the rotor speed by checking the REV MIN x 1000 indicator A zero speed switch is incorporated in the REV MIN x 1000...

Page 17: ...erate Wait until the rotor stops spinning before using the mechanical override The mechanical override loop is located under the right front corner of the centrifuge as shown in figure 3 2 To operate...

Page 18: ...uired to prevent rotor failure Observe WARNING on the Safety Information Page in front of this manual The rotor speed is reduced in proportion to the square of the ratio for the maximum allowable comp...

Page 19: ...tronics should test or repair the electrical circuits If hazardous materials have been processed in the centrifuge take all necessary precautions when cleaning or servicing the centrifuge to avoid per...

Page 20: ...polish occasionally with 400 Emery cloth Always wipe the spindle clean after polishing Cabinet Clean the enameled cabinet panels with a household wax cleaner Use a mild nonalkaline detergent and water...

Page 21: ...b If the brushes are shorter than 6 mm 1 4 inch replace worn brushes with new brushes Catalog No 12732 NOTE If this is not the original set of brushes and the brushes are shorter than 6 mm 1 4 inch pe...

Page 22: ...cing resulting in excessive brush and commutator wear Figure 4 2 Motor Brush Assembly CHAMBER TEMP C Indicator Adjustment RT6000B only With the POWER switch set to 0 use a screwdriver and adjust the a...

Page 23: ...ruction manual contains specific guidance about cleaning and decontamination methods appropriate for the product it describes Clean and decontaminate your centrifuge or rotor as follows For tabletop c...

Page 24: ...e or rotor at its Service facilities which in its opinion is a radioactive or biological hazard the sender will be contacted for instructions as to disposition of the equipment Disposition costs will...

Page 25: ...SORVALL Centrifuges Maintenance 4 7...

Page 26: ...APPENDIX...

Page 27: ...he plate affixed to the instrument or use of any rotor other than those intended for use in this instrument Du Pont reserves the right to change alter modify or improve any of its instruments without...

Page 28: ...fuge 2 2 Installation Rotor 3 3 Location Centrifuge 2 2 Maintenance 4 1 Mass Centrifuge 1 2 Mechanical Override Loop 3 6 Motor Brush Replacement 4 3 Ordering Information 4 4 Operating Temperature RT60...

Reviews: