Hybrid Recorder mode starts when a Start Record signal is received and
continues until the unit receives a trigger signal and then records for the
post-trigger time specified by the host software. The trigger signal marks
the T=0 point and all data recorded is available for download.
5.2.2.4.
Continuous Recorder Mode
Continuous Recorder mode starts when a Start Record signal is received
and continues until the Start Record signal is released. The unit will then
re-arm for another event. The LEDs on the unit will flash blue slowly then
rapidly, and then the STATUS LED will become solid blue, indicating the
unit is fully armed. The unit will continue to record new events until it
records the number of events specified by the host software. If a trigger
signal is received after the unit has re-armed the unit will disarm and no
longer attempt to re-arm.
5.2.3.
Multiple-Event Modes
All SLICE data collection modes have an equivalent multi-event arming mode. A
unit armed in a multiple-event mode will re-arm when an event completes. The
unit will stop re-arming when the number of events specified by the host
software has been recorded.
5.2.4.
Auto-Arm Data Collection
SLICE can be placed in an auto-arm mode that will cause the unit to arm
automatically when the power cycled. The unit can be placed into this mode
and record with any data collection mode.
NOTE:
An event or trigger signal applied anywhere in the SLICE chain is
distributed throughout the system. This applies to level trigger as
well.