No  special  knowledge  or  training  is  required  to  operate  this  equipment.  Please  read  the  instructions 
carefully before use. 

 

Keep these instructions in a safe place with the appliance as it contains the necessary information for the 
proper use of the equipment. 

 

This medical device is to be used either in health care establishment or at home. 

 

The user can safely use all the functions of the breast pump. 

 

For hygiene reasons, a new complete KIT EXPRESSION must be sold to each mother. 

 

WARNINGS: 

 

Do not touch the breast pump box for more than 1 second during use. 

 

The tube connecting the collection container to the breast pump can reach the temperature of 43.5°C. The 
user should not touch the tube for more than 10 minutes during the breast pump use. 

 

Notes on the functionality of the breast pump 

 

The international version of the KOLOR breast shield (with an international screw thread cap) can only be 
used with a matching collection container (same screw thread cap). 

 

Switch off the breast pump when the collection container is filled in order to prevent milk from getting into 
the interface bottle. 

 

If the device has to be rinsed, entrust the operation to a maintenance professional (risk of bad handling). 

 

CAUTION, in all cases: 

 

• WARNING: Always disconnect the plug from the power supply before opening it. 
• WARNING: Only an approved customer service is authorized to repair the breast pump. 
• WARNING: Modifications are forbidden. 
• WARNING: Only use the accessories supplied and mentioned in the instructions for use. 
• WARNING: The warranty does not cover connection errors with non-compatible accessories. 
• WARNING: Do not undertake service or maintenance operations while the breast pump is used. 

 

7.

 

Problem solving 

 

Problem 

Solution 

No operating noise 

can be heard from 

the pump 

Make sure that the power cord is connected to the power supply. 
Check whether the switch is on the operating position. 
Check that the current delivered by the power supply corresponds to that of the breast pump (Volt 
Hz). 
Check with another breast pump if power supply is available in the wall outlet. 
If the problem continues, please contact DTF MEDICAL or your local distributor. 

There is low or no 

suction 

Check that the breast shield is fully screwed onto the collection container (screw tightly). 
Check that every tube is properly connected and not pinched. 
Check that the cap of interface bottle correctly fits. 
Check the condition of the O-ring between the breast shield and the collection container. 
Check that there is no residue in the tube. 
If the problem continues, please contact DTF MEDICAL or your local distributor. 

The suction is too 

important 

Switch off then switch on the breast pump. 
If the problem continues, please contact DTF MEDICAL or your local distributor. 

The breast pump 

stops abruptly when 

used 

Disconnect the tube from the interface bottle to vent it. 
In case of alternating stop, please contact DTF MEDICAL or your local distributor. 

The pump sucks 

continuously 

Do not tear off the breast shield from your breast (could cause pain and/or damage). 
Insert  your  finger  between  your  breast  and  the  breast  shield  to  let  the  air  come  in  and  remove  it 
more easily or disconnect the tube from the breast shield. 

A piece or an 

accessory is missing 

Please contact DTF MEDICAL or your local distributor. 

The breast pump is 

too noisy or 

malfunctioning 

Please contact DTF MEDICAL or your local distributor. 

Milk entered inside 
the interface bottle 

Clean it as soon as it has been soiled or discard it if you are in a healthcare establishment. 
Check if the breast pump is not wet or damp. If it is, do not use the breast pump. 
Do not use the milk that penetrated into the interface bottle or tubes. 
The tubes should be briefly dipped into boiling water and then thoroughly dried. 

Summary of Contents for kitett FISIO PRO

Page 1: ...VI H ng d n s d ng 0459 2003 LA DIFFUSION TECHNIQUE FRANCAISE 19 rue de la Presse CS 60132 42003 SAINT ETIENNE Cedex 1 France Phone number 33 0 4 77 74 51 11 Fax 33 0 4 77 79 67 72 Email dtf dtf fr w...

Page 2: ...use 3 Electric breast pump associated with KOLOR KIT EXPRESSION TR Kullanma k lavuzu 14 Elektrikli g s pompas KOLOR KIT EXPRESSION ile ili kilidir VI H ng d n s d ng 25 M y h t s a i n li n k t v i K...

Page 3: ...ning at home 5 c Disinfection at home 6 d Cleaning and sterilization by healthcare establishments 6 4 Use 7 a Breast pump functioning 7 b Accessories preparation 7 c Using the breast pump 7 d After ea...

Page 4: ...e used in single or double pumping and should be plugged into to the power supply It combines a suction power adjustment with a set up rhythm The rhythm and the suction power are simultaneously adjust...

Page 5: ...select the correct inner diameter size The nipple should not feel too constricted neither too free You should see space around your nipple in order to let it move easily in the breast shield during pu...

Page 6: ...on a paper towel or clean linen NOTES In the case of intensive disinfection premature infants or with a weakened immune system boiling water is the only method that guarantees an optimal lifetime of...

Page 7: ...of the double KIT EXPRESSION can also be used as single one WARNING Correctly close the interface bottle cap when the breast pump is used on simple pumping c Using the breast pump Place the KOLOR bre...

Page 8: ...network The warranty does not cover faulty connections This medical device is designed to operate at altitudes less than or equal to 3 000 meters Do not open the breast pump when connected to the powe...

Page 9: ...that the power cord is connected to the power supply Check whether the switch is on the operating position Check that the current delivered by the power supply corresponds to that of the breast pump...

Page 10: ...Plastic interface bottle for double or single pumping W2 Soft transport bag for the breast pump MBOXK Foam filter YR36 KIT EXPRESSION Kit Ref Breast shield size Shape size Single Double Kit Ref Breast...

Page 11: ...A Fuses T 0 5A H 250V Suction values Variable from 90 to 300hPa Cycles 40 to 140 cycles minutes Sound level 57dBa Running mode Continuous Power supply Alternative current 230V 50Hz Ambient temperature...

Page 12: ...11 Group 1 RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacturer s declaration electromagnetic...

Page 13: ...se Warning Dangerous voltage Caution Warning Keep dry Fear of moisture Fragile Handle with Care Store away from light This symbol indicates the limit of atmospheric pressure not to be exceeded during...

Page 14: ...ar 16 c Evde dezenfeksiyon 17 d Sa l k kurumlar taraf ndan temizlik ve sterilizasyon 17 4 Kullan m 18 a G s pompas fonksiyonu 18 b Aksuarlar n haz rlanmas 18 c G s pompas n kullanma 18 d Her kullan m...

Page 15: ...ya ama sahip anneler i in idealdir FISIO pompa tekli olarak ya da iki g se de ayn anda tak lacak ekilde kullan labilir ve bir g kayna na ba lanmal d r Emme g c ayar n ritim ayarlama ile birle tirmekte...

Page 16: ...i erisinde kolayca hareket etmesi i in meme ucunuzun evresinde bo luk g rmeniz gerekir E er emme ba l ok s karsa meme ucunuzda a r yara veya kanallar n tak nmas na neden olacakt r E er emme ba l ok g...

Page 17: ...ay n NOTLAR Kapsaml dezenfeksiyon durumunda erken do an ya da zay f ba kl k sistemine sahip bebekler i in aksesuarlar n optimum mr n garanti eden en iyi y ntemdir Dezenfeksiyon kaynar suda yap l rsa v...

Page 18: ...aray z biberonuna ba lay n Tekli KIT EXPRESSION kurulumu kili KIT EXPRESSION kurulumu tekli olarak da kullan labilir UYARI G s pompas tekli pompalama ile kullan ld nda aray z biberonunun kapa n do ru...

Page 19: ...nt durumunda garanti uygulanmamaktad r Bu t bbi cihaz 3000 metre veya alt ndaki rak m seviyelerinde al mas i in tasarlanm t r G kayna na ba l oldu u durumlarda g s pompas n n i ini a may n Uyar Bu t b...

Page 20: ...gelmiyor G kablosunun g kayna na ba l oldu undan emin olun Anahtar n al ma konumunda olup olmad n kontrol edin G kayna taraf ndan sa lanan ak m n g s pompas na uygun oldu unu kontrol edin Volt Hz Priz...

Page 21: ...ir KIT EXPRESSION i in 1 metre uzunlu unda silikon t p TU32 kili ya da tekli pompalar i in plastik aray z biberonu W2 G s pompas i in yumu ak ta ma antas MBOXK K p k filtresi YR36 KIT EXPRESSION Kit R...

Page 22: ...220V 60Hz s r mleri mevcuttur T ketim 100VA Sigortalar T 0 5A H 250V Emme de erleri 90 ile 300hPa aras de i iklik g stermektedir Devirler 40 ile 140 aras devir dakika Ses d zeyi 57dBa al ma modu S rek...

Page 23: ...nlar CISPR 11 S n f B Harmonik emisyonlar IEC 61000 3 2 S n f A Voltaj dalgalanmalar titre im emisyonlar IEC 61000 3 3 Uyumludur Rehber ve reticinin beyan elektromanyetik ba kl k Ba kl k testi IEC 606...

Page 24: ...nma k lavuzuna uyunuz Uyar Tehlikeli voltaj Dikkat Uyar Kuru tutun Nem riski Hassas Dikkatli kullan n G ne ndan uzakta saklay n Bu sembol al ma ta ma veya saklama s ras nda ge ilmemesi gereken atmosfe...

Page 25: ...n t i nh 27 c Kh tr ng t i nh 28 d V sinh v kh tr ng b i c s ch m s c s c kh e 28 4 S d ng 29 a V n h nh m y h t s a 29 b Chu n b ph ki n 29 c S d ng m y h t s a 29 d Sau m i l n s d ng 30 5 Cho con...

Page 26: ...con b ho c cho con b tr c ti p kh ng V i t nh n ng d d ng s d ng y l m y l t ng cho c c b m n ng ng C th s d ng m y h t s a FISIO ch h t s a n ho c i v c n c m m y v o ngu n i n M y k t h p gi a i u...

Page 27: ...anh n m v ch n ng k ch th c ng k nh trong Kh ng n n n m v qu ch t c ng nh qu l ng N n m t kho ng kh ng quanh n m v di chuy n d d ng trong ph u ch p h t khi h t s a Ph u ch p h t qu ch t c th g y au t...

Page 28: ...1000 Watt kh t nhi n tr n kh n gi y ho c v i s ch L U Trong tr ng h p kh tr ng chuy n s u tr sinh non ho c c h mi n d ch y u n c s i l ph ng ph p duy nh t m b o tu i th t i u c a ph ki n N u kh tr ng...

Page 29: ...a K t n i ng h t v i u v i v o c a ph u ch p h t v v i b nh trung gian tr n m y Thi t l p b KIT EXPRESSION n Thi t l p b KIT EXPRESSION i b n y c ng c th c s d ng nh b n C NH B O ng ch nh x c n p b n...

Page 30: ...tr xu ng Kh ng m m y h t s a khi k t n i v i ngu n i n C nh b o thi t b y t n y c k t n i tr c ti p v i ngu n i n kh ng s d ng trong b n t m ho c d i v i hoa sen kh ng ng m ch m Gi thi t b y t n y tr...

Page 31: ...i DTF MEDICAL ho c nh cung c p t i a ph ng L c h t y u ho c kh ng h t Ki m tra xem ph u ch p h t c v t h t c v o b nh ng s a hay ch a v n ch t Ki m tra xem c c ng c c n i ph h p v kh ng b k p hay kh n...

Page 32: ...ch th c h nh d ng n i K21ES 21mm Small n K24ES 24mm Small n K21EL 21mm Large n K24EL 24mm Large n K21ESD 21mm Small i K24ESD 24mm Small i K21ELD 21mm Large i K24ELD 24mm Large i K26ES 26mm Small n K30...

Page 33: ...p C s n 230V 50Hz 220V 50Hz ho c 220V 60Hz M c ti u th 100VA Ng t m ch T 0 5A H 250V Gi tr h t Bi n thi n t 90 n 300hPa Chu k 40 n 140 chu k ph t M c m thanh 57dBa Ch ch y Li n t c Ngu n i n D ng i n...

Page 34: ...ao ng i n p ph t x nh p nh y IEC 61000 3 3 Tu n th H ng d n v tuy n b c a nh s n xu t mi n nhi u i n t Ki m tra mi n nhi u M c ki m tra IEC 60601 M c tu n th Ph ng t nh i n ESD IEC 61000 4 2 8 kV khi...

Page 35: ...uy hi m Th n tr ng C nh b o Gi kh ph ng m t D v X l c n th n B o qu n tr nh nh n ng Bi u t ng n y cho bi t kh ng c v t qu gi i h n p su t kh ng kh trong khi v n h nh v n chuy n ho c b o qu n Bi u t ng...

Reviews: