6

2.1.13 After finishing operation, keep the machine working for 30 seconds with the water supply to clean the handpiece and the tip.

2.1.14 Unscrew the scaling tip  and sterilize it.

2.1.15 After operating, turn off the power switch to cut off the power, then pull off the plug of power adapter.

Note: Don't screw the scaling tips when stepping on the foot switch, and the machine is working.

3. Maintenance and sterilization

3.1 Sterilization of detachable handpiece

3.1.1 Autoclaved under high temperature, pressure, time: 

a) 134°C, 2.0bar~2.3bar (0.20MPa~0.23MPa), 4min.

b) Pull out the handpiece and unscrew the scaling tip and endochuck after each operation.

c) Pack the handpiece with sterile gauze or sterile bag before sterilizing.

d) Reuse handpiece after it cools naturally to avoid scalding hands.

3.1.2 Notice

a) Dry the cleaning liquid in the handpiece with compressed air before sterilization. 

b) Be sure that the scaling tip has been unscrewed from the handpiece and it cannot be sterilized with others.

c) Please notice whether the outer of the handpiece is damaged during the treatment or sterilization, don't smear any protective oil on the surface 

of handpiece.

d) There are two waterproof "o" rings at the end of handpiece. Please lubricate them with dental lube frequently, as sterilization and repeated 

pulling and inserting will reduce their service life. Change a new one once it is damaged or worn excessively. 

e) The following sterilizing methods are forbidden:

i. Put handpiece into any liquid for boiling.

ii. Dip handpiece in disinfectors such as iodine, alcohol and glutaraldehyde.

iii. Put handpiece into oven or microwave oven for baking.

3.2 Sterilization of tip and Endochuck

3.2.1 Select the temperature, pressure and time of the autoclave sterilizer:

         134°C,2.0bar-2.3bar (0.20MPa-0.23MPa)

4min

.

3.2.2 Please remove tip and Endochuck after each operation

3.2.3 The surgical tip and Endochuck should be individually wrapped and sterilized with a sterilization towel or bag.

3.2.4 After sterilization, it is necessary to cool the tip before it can be used again so as not to burn the hands.

3.2.5 Before sterilization, please remove the tip from handpiece. Don’t sterilize tip with other instrument in the same sterilizer.

Summary of Contents for S6 LED

Page 1: ...S6 LED ULTRASONIC SCALER INSTRUCTION MANUAL...

Page 2: ...llation and adjustment 5 3 Maintenance and sterilization 6 4 Precaution 10 5 After service 12 6 Symbol instruction 12 7 Environmental protection 13 8 Manufacturer s right 13 9 European authorized repr...

Page 3: ...nce and structural composition Ultrasonic scaler S6 LED is composed of electrocircuit water way and ultrasonic transducer 1 2 3 Scope of application Ultrasonic scaler S6 LED is used for the dental cal...

Page 4: ...itrous Oxide Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 3 17 The main unit use a power adapter to connet network power s...

Page 5: ...3 1 4 2 Back of the main unit sketch map Picture 2 1 4 3 Instruction of using wrench Picture 3...

Page 6: ...4 Picture 4 1 4 4 List of removable parts Picture 5 Picture 6...

Page 7: ...ee picture 4 2 1 6 Turn on the power switch the power indicator lighted and the machine is ready for work Touch panel is applied to this machine power can be adjusted by directly touching the power in...

Page 8: ...notice whether the outer of the handpiece is damaged during the treatment or sterilization don t smear any protective oil on the surface of handpiece d There are two waterproof o rings at the end of...

Page 9: ...while touching teeth 3 5 Cleaning of tips endochuck torque wrench and endo wrench The scaling tip endochuck torque wrench and endo wrench can be cleaned by ultrasonic cleaner 3 6 Troubleshooting and n...

Page 10: ...er There is impurity in the solenoid valve Contact with the local distributor or manufacturer The handpiece generates heat The amount of spouting water is too little Turn the water control switch to a...

Page 11: ...2 Notice Note 1 The water control knob can adjust the water volume according to the symbol Note 2 To clean the water pipe with the multi function syringe of the dental unit see picture 7 Multi functi...

Page 12: ...g out Please keep the water flow smoothly 4 1 9 Don t use impure water source and be sure not to use normal saline instead of pure water source 4 1 10 If use the water source without hydraulic pressur...

Page 13: ...a room where the relative humidity is 10 93 atmospheric pressure is 70kPa to106kPa and the temperature is 20 C 55 C 4 3 4 Please turn off the power switch and pull out the power plug when the equipme...

Page 14: ...Symbol instruction Trademark Ordinary equipment Alternating current Drip proof Date of manufacture Manufacturer Class II equipment Type B applied part Foot switch interface Used indoor only Adjustment...

Page 15: ...the electromagnetic environment specified below The customer or the user of the models S6 LED should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Page 16: ...rruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec...

Page 17: ...ansmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equipment marked with the follow...

Page 18: ...2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Page 19: ...ZMN WI 09 416 V1 2 20181102...

Reviews: