background image

10

[Note 3 ] If the scaling tip has been screwed on tightly and there is fine spray too, the following phenomena show that the scaling tip is damaged:

a) The vibrating intensity and the pulverization degree become weak obviously.

b) During operating, there is some buzz when the scaling tip is working.

[Note 4] If the power cable connected to the grid electricity end of the power adapter is damaged, as the power cable is not removable, for your 

safety, please contact a professional for replacement.

4. Precaution

4.1 Usage notice

4.1.1The use of the product must comply with the relevant regulatory requirements  of the medical regulatory authorities, and should only be 

used by trained doctors or technicians.

4.1.2 Keep the scaler clean before and after operation.

4.1.3 The scaling tip, wrench and handpiece must be sterilized before each treatment.

4.1.4 Don't screw the scaling tip when stepping on the foot switch.

4.1.5 The scaling tip must be fastened. There must be fine spray coming out from the tip when operating.

4.1.6 Change a new one when the tip is damaged or worn excessively.

4.1.7 Don't twist or rub the tip.

4.1.8 While scaler working, the heat of scaling tip will become higher if there is no water flowing out. Please keep the water flow smoothly. 

4.1.9 Don't use impure water source, and be sure not to use normal saline instead of pure water source.

4.1.10 If use the water source without hydraulic pressure, the water surface should be one meter higher than the head of the patient.

4.1.11 Don't knock or rub the handpiece.

4.1.12 Please put the power plug into the socket easy to pull out, to make sure it can be pull out in emergency.

4.1.13 When using the equipment, please keep the water get through smoothly, otherwise patient's tooth surface would be injured by overheat in 

the handpiece.

4.1.14 After operating, turn off electrical source, and then pull out the plug of power adapter.

4.1.15 As a professional manufacturer of medical instruments, we are only responsible for the safety on the following conditions:

I.The maintenance, repair and modification are made by the manufacturer or the authorized dealer.

II. The changed components are original of "DTE" and operated correctly according to instruction manual.

III. Maintenance is forbidden while the device is used for treatment.

4.1.16 The screw thread of the scaling tips produced by other manufacturers maybe coarse, rusty and collapsed, which will damage the screw 

thread of the handpiece irretrievably. Please use "WOODPECKER" brand scaling tip.

4.1.17 Please select a suitable power when using different type of tips (refer to "TABLE OF OPERATING POWER OF THE TIPS").

Summary of Contents for S6 LED

Page 1: ...S6 LED ULTRASONIC SCALER INSTRUCTION MANUAL...

Page 2: ...llation and adjustment 5 3 Maintenance and sterilization 6 4 Precaution 10 5 After service 12 6 Symbol instruction 12 7 Environmental protection 13 8 Manufacturer s right 13 9 European authorized repr...

Page 3: ...nce and structural composition Ultrasonic scaler S6 LED is composed of electrocircuit water way and ultrasonic transducer 1 2 3 Scope of application Ultrasonic scaler S6 LED is used for the dental cal...

Page 4: ...itrous Oxide Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 3 17 The main unit use a power adapter to connet network power s...

Page 5: ...3 1 4 2 Back of the main unit sketch map Picture 2 1 4 3 Instruction of using wrench Picture 3...

Page 6: ...4 Picture 4 1 4 4 List of removable parts Picture 5 Picture 6...

Page 7: ...ee picture 4 2 1 6 Turn on the power switch the power indicator lighted and the machine is ready for work Touch panel is applied to this machine power can be adjusted by directly touching the power in...

Page 8: ...notice whether the outer of the handpiece is damaged during the treatment or sterilization don t smear any protective oil on the surface of handpiece d There are two waterproof o rings at the end of...

Page 9: ...while touching teeth 3 5 Cleaning of tips endochuck torque wrench and endo wrench The scaling tip endochuck torque wrench and endo wrench can be cleaned by ultrasonic cleaner 3 6 Troubleshooting and n...

Page 10: ...er There is impurity in the solenoid valve Contact with the local distributor or manufacturer The handpiece generates heat The amount of spouting water is too little Turn the water control switch to a...

Page 11: ...2 Notice Note 1 The water control knob can adjust the water volume according to the symbol Note 2 To clean the water pipe with the multi function syringe of the dental unit see picture 7 Multi functi...

Page 12: ...g out Please keep the water flow smoothly 4 1 9 Don t use impure water source and be sure not to use normal saline instead of pure water source 4 1 10 If use the water source without hydraulic pressur...

Page 13: ...a room where the relative humidity is 10 93 atmospheric pressure is 70kPa to106kPa and the temperature is 20 C 55 C 4 3 4 Please turn off the power switch and pull out the power plug when the equipme...

Page 14: ...Symbol instruction Trademark Ordinary equipment Alternating current Drip proof Date of manufacture Manufacturer Class II equipment Type B applied part Foot switch interface Used indoor only Adjustment...

Page 15: ...the electromagnetic environment specified below The customer or the user of the models S6 LED should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Page 16: ...rruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec...

Page 17: ...ansmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equipment marked with the follow...

Page 18: ...2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Page 19: ...ZMN WI 09 416 V1 2 20181102...

Reviews: