5 -
microwaves. This condition may affect the result.
⑤
Check Location of Installation.
Install the reader in a place with humidity less than 90%.
Install the reader in a well-ventilated place with minimal dust.
Avoid any place where physical and chemical effect is possible.
Avoid direct sunlight.
Install the reader on a flat surface.
If room temperature is too high, an error may arise.
(Optimal operating temperature: 10~40
℃
)
The use of this reader with explosive, inflammable or putrefactive substances
is prohibited.
The use of this reader in a place with radiation, a magnetic field or a vacuum
is prohibited.
The use of this reader in a place with extreme electric noise is prohibited.
The Calibrator may be discolored under direct sunlight. It is recommended that
the calibrator is stored in a dark place.
Be careful not to extremely tilt or drop the reader during transfer.
Notice
This product optically analyzes the reaction between specific pathogenic organism and
reagent in the diagnostic process of specimen, and quantifies degree of the reaction
This system is for quantitative analysis of lateral flow immunochromatographic devices. The
system can be operated by a person of a non-medical field, but the result should be
interpreted and analyzed by medical professionals.
You can buy supplies like the calibrator and test devices from your supplier. Supplies should
be tested for adaptability with the system before it is actually used for the test. Information
about applicable supplies is available by contacting DSM.