AN401Plus 

QUICK START MANUAL

 

 

 

Front panel 

 

Rear panel  

 

Power supply connection

 

= AN401 Power supply 

= AN401 Power supply 
 

Polarity free  

 

Slot C Transducer connections:

 

Terminal 

Differential signal input card [I1 option] 

Single ended input card [I2 option] 

 

Transducer’ Power supply – (GND) 

Signal - 

+ Input signal (0 to + 5 or 10V) 

Not connected 

Not connected 

Not connected 

Not connected 

Transducer’s Power supply - 

Not connected 

Transducer’s Power  

Transducer’s Power  

1.

 

Panel cut

: 92 x 44 mm, depth: 110 mm (including rear terminal board).  

2.

 

Power supply

:

 connect power supply to the instrument by means of the screw terminal 

, without bothering about polarity (0 to 

8 - 24 Vdc and 0 to 9 - 15 Vac, 1W power ). 

3.

 

Transducer connection

:

 See Label on top of the instrument

 

4.

 

Calibration

:

  

A.

 

Power on the AN401Plus and push, at the same time, button 

and 

C

; when display shows 

FSET

;

 

release push buttons (by 

releasing push buttons You enter into calibration menu).   Wait at least 10-15min for warm up time. 

B.

 

Display shows 

LO

: press 

E  

push button

 

to store the ZERO numeric value (in the desired engineering units i.e. Kg, tons, daN 

etc.).      No tare or pre-load is applied to the transducer (i.e.: 0 Kg = 

0 default value

 ).    

Refer to point E about how to set different ZERO value. 

C.

 

Verify that load cell is free from any load and then press 

E

  (

MEMO 

blinks on display

)

D.

 

Display shows 

LO

: press 

 

in order to show

 

HI

 . 

E.

 

Display shows 

HI

: press 

and write the FULL SCALE numeric value (in the desired engineering units i.e. Kg, tons, daN etc.), 

corresponding to a specimen applied onto the load cell or by using built-in calibration circuit: i.e. 500 Kg (

1000 = default

).  

To increase or decrease the value press 

 

or

; to increase It quickly press 

 

and 

 

E

, at the same time, or 

and

 

E

 to decrease 

It quickly. 

F.

 

Apply the specimen onto the load cell, or activate the built-in calibration circuit, and press  

to memorize It: on display It 

blinks 

MEMO

G.

 

Display shows 

HI

:

 

press 

 in order to show

 

dp

 . 

H.

 

Display shows 

dp

  (Decimal Point): press 

to set the decimal point position.  

In example, to show 100,0 , set value to 2 (decimal point is set on second digit from right side) and press 

E

I.

 

Display shows 

dp

: press 

 

to show 

adspd

 (ad speed = analog/digital sampling rate). 

Default value is 8=15 Hz

. Values are: 0 = 

3.820 kHz, 1 = 1.910 kHz, 2 = 960 Hz, 3 = 480 Hz, 4 = 240 Hz, 5 = 120 Hz, 6 = 60 Hz, 7 = 30 Hz, 8 = 15 Hz e 7 = 7.5 Hz. 
Resolution depends on sampling rate: if high speed measures are not necessary, do not change default value. 

J.

 

The AN401Plus is calibrated: press 

C

 to exit from calibration and personalization menu (AN401Plus restarts Itself and display 

shows the transducer’s measure; 

default

 = MODO 1, NET value). 

5.

 

Dynamic zero

: after calibration, the AN401Plus can show the transducer measures and, by pressing 

C

, It will zeroes the displayed 

measured value = DYNAMIC ZERO (this zeroing is not maintained after powering down the AN401Plus). 

6.

 

Grounding

: if heavy electrical environmental noise is present the grounding pin (earth), on the rear panel of AN401Plus, is to be 

connected to a good grounding well. 

7.

 

Alarm levels menu

:  by pressing 

 

It is possible to set alarm levels values (see rear page). 

8.

 

Complete manual

: for a comprehensive list of AN401Plus functions make reference to the complete manual available under request 

or from web site 

www.dseurope.com

 . 

 

1

2

3

4

1

3

4
2

5 4 3 2 1

9 8 7 6

CanBus

RS-485

RS-232

INPUT

1

2

3

4

5

1
2
3
4
5
6

6

4
5

3

1
2

6

4
5

3

1
2

6

4
5

3

1
2

D

C

B

A

EARTH

RJ-11A

RJ-11B