![Drucker Diagnostics HORIZON 24-AH Operator'S Manual Download Page 13](http://html1.mh-extra.com/html/drucker-diagnostics/horizon-24-ah/horizon-24-ah_operators-manual_2533138013.webp)
Drucker Diagnostics | +1-814-692-7761 |
| 11
G
ENERAL
S
PECIFICATIONS
The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Tube Capacity
24 tubes (3-6 mL)
6 tubes up to 60 mL
Dimensions (H x W x D)
15 in x 17 in x 9 in
(38 cm x 43 cm x 23 cm)
Weight
39 lbs. (17 kg)
Sound Level
59 dB A
Environmental Range
16 – 32 °C
Voltage
95 -253 VAC
Frequency
50/60 Hz
Power Requirement
280 Watts
Centrifuge Motor
½ H.P. Brushless DC
Cycle Time (in Minutes)
1-30 (+/- 2%)
Max Speed
2,500 RPM
C
ALCULATING THE
G-F
ORCE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to
www.druckerdiagnostics.com/g-force-calculator/
.
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
2
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
2
Radius
6 in. (15.3 cm)
Summary of Contents for HORIZON 24-AH