Drucker Diagnostics DASH Flex Series Operator'S Manual Download Page 11

Drucker Diagnostics 

 Customer Service: +1-814-692-7661 

 [email protected] | Page 7  

 

D

ELETING A 

C

YCLE

 

 

Enter the  

Advanced Menu  

With the desired cycle selected, access the menu and enter the 
advanced menu. 

 

Navigate to Delete 

Using the UP and DOWN buttons, navigate to DELETE. Exit the 
menu. 

WARNING: CYCLE WILL BE DELETED IF MENU IS EXITED 

WITH DELETE SELECTED 

 

 

Confirm Deletion 

Press the CYCLE button to Delete the cycle

 

 

C

YCLE 

L

OCK

 

To ensure repeatability, the centrifuge can be locked either on one cycle (Single Cycle Lock) or restricted to the saved 
cycles (Preset Lock). The Single Cycle Lock also prevents making changes to the selected cycle parameters. The Preset 
Lock allows selection of any saved cycle and prevents changing the parameters of saved cycles. 
 

 

Enter Single Cycle 

Select desired cycle. With lid open, press and hold the UNLOCK 
button. One beep will confirm that cycle selection is locked. 

 

Enter Preset Lock 

Continue holding the UNLOCK button to enter Preset Lock. Two 
beeps will confirm that cycle selection is now locked. 

NOTE: If 

Single Cycle Lock is set, it must be canceled before Preset Lock 
can be set.

 

 

Cancel Lock 

Hold the UNLOCK button. Three beeps will confirm that the cycle 
selection is now unlocked. 

 

 

P

RESET 

C

YCLES

 

 

 

(1) Chemistry 

(2) Coag (PPP) 

(3) Urine 

HORIZON 12 

Flex 

RPM 

3,500 

3,200 

1,800 

Time 

10 

15 

G-Force 

1,800 xg 

1,500 xg 

500 xg 

 

 

 

 

 

HORIZON 24 

Flex 

RPM 

3,300 

3,000 

1,700 

Time 

10 

15 

G-Force 

1,800 xg 

1,500 xg 

500 xg 

Summary of Contents for DASH Flex Series

Page 1: ...HORIZON 12 Flex 24 Flex Operator s Manual...

Page 2: ......

Page 3: ...Menu 4 Free Mode 5 Create New Cycle 5 Display Cycle Count 5 Modifying a Preset 6 Deleting a Cycle 7 Cycle Lock 7 Preset Cycles 7 Operation 8 Balancing Loads 8 12 Tube Centrifuges 8 24 Tube Centrifuge...

Page 4: ......

Page 5: ...being left in the centrifuge longer than necessary patent pending o A traditional audible alert indicates the completion of the cycle The audible alert can be muted o Cool Flow design prevents overhe...

Page 6: ...is operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be required to correct the interf...

Page 7: ...ry preset for Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycl...

Page 8: ...d the CYCLE button until you hear a double beep The cycle number should begin flashing Enter the Advanced Menu Press the GEAR button to enter the advanced menu FROM FREE MODE Access the Menu Press the...

Page 9: ...IME display Change Brake Values Access the menu Press the down key next to the top screen until it shows Brake To turn brake on and off use the up and down buttons next to the TIME display Exit the me...

Page 10: ...CF button Setting Time Press the UP and DOWN buttons next to the TIME display Change Brake Values While in the advanced menu navigate to Brake Use the UP and DOWN buttons next to the TIME screen to tu...

Page 11: ...changes to the selected cycle parameters The Preset Lock allows selection of any saved cycle and prevents changing the parameters of saved cycles Enter Single Cycle Select desired cycle With lid open...

Page 12: ...ng mechanism will engage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed press the UNLOCK button The lid will unlock for another 15 seconds o Tur...

Page 13: ...must be loaded symmetrically with tubes 15 mL 125 mm tubes may be spun in the standard HORIZON 24 Flex buckets but must be loaded as follows When three buckets are being used each bucket must be loade...

Page 14: ...or storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a...

Page 15: ...n slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise or vibration when the centrifuge is running o Verify tha...

Page 16: ...annot be rotated press the UNLOCK button and try again o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use a pen to manuall...

Page 17: ...l Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 280 Watts 280 Watts Centrifuge Motor H P Brushless H P Brushless Max g Force 2 000 xg...

Page 18: ...cling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is r...

Reviews: