background image

Copyright 2022 Baker Hughes Company.

2 | UNIK5600/5700 Demountable Backend Exchange Manual–English

3.3  Re-assembly

1. Exchange O-ring and put it into the seal groove. Make sure the O-ring is fully seated in the 

groove. Please always replace with a new O-ring when reassembling the backend. The part 

number of the O-ring is 163M2956-1.

2. Reconnect the connector, ensuring correct alignment. Please pull back slightly to make sure 

it is not pulled out and fully fixed.

3. Assemble the screws and spring washers. Ensure screws and washers are undamaged 

before installing.