1.  INTRODUCTION 

This manual contains the removal and installation procedure for the backend of the 

demountable KYNAR cable and demountable TPE-U cable version of the UNIK5600/5700.

Read the whole installation procedure before starting installation. The technical specification 

is detailed in the datasheet.

2.  CAUTIONS

All the operations described in this manual should only be performed by trained and 

authorized professionals.

Always wear the necessary personal protective equipment when completing this procedure.

3.  INSTALLATION

3.1  Accessories & Tools Preparation

Torque wrench (1.8 Nm)

2.5mm Hexagon bit

O-ring (163M2956-1)

3.2  Disassembly Procedure

1. Remove the screws and spring washers.
2. Separate the flange and unplug the connector. Make sure the flying leads are not pulled out 

from the connector.

WARNING  

Do not complete this procedure in a hazardous area while an 

explosive atmosphere is present - THIS CAN CAUSE AN EXPLOSION. Isolate the 

power supply to the equipment first.