K0463 Revision A - [EN] English
S2 PV
622-IS models: Start operations
S2.1
Release the pressure/Attach the device under test
S2.2
Vacuum or pressure operation
A2.2 (Front cover)
1
Step
Procedure
1.
Release the pressure: Open the refill
valve (1 turn) then fully open the
pressure release valve.
2.
Use the applicable adaptor to attach
the device; figure A3.
4
5
2
2
a
b
7
6
Step
Procedure (Vacuum)
1.
Set to vacuum operation (-).
2.
Open the refill valve (1 turn).
3.
Wind the volume adjuster to
mid-range or fully-clockwise.
4.
Seal the system.
5.
Set the vacuum with the pump.
6.
Adjust the vacuum
(+ decrease; - increase).
Step
Procedure (Pressure)
1.
Set to pressure operation (+).
2.
Wind the volume adjuster to
mid-range.
3.
Seal the system.
4.
Use the pump to set a pressure up to
20 bar (300 psi).
5.
Open the refill valve (1 turn) to have full
control to increase (+) or decrease (-)
the pressure with the volume adjuster.
6.
If increasing pressure and get to the
limit of travel, close the refill valve and
wind the volume adjuster fully
counter-clockwise. There is no change
in pressure.
7.
Refill the pressure mechanism with the
pump (
15 cycles) and wind the
volume adjuster clockwise until the
pressure starts to increase.
8.
Continue to do steps 5 to 7 until the
necessary pressure is achieved.
11
5
9
4
12
9
11
9
4
12
9
5
9
5
9
12
5
+
-
6
1
4
3
5
2
1
2
3
4
+
-
+
Summary of Contents for PV 62X-IS
Page 2: ......
Page 6: ...Copyright 2009 Baker Hughes Company iv PV 62X IS Quick Start Guide English ...
Page 18: ...12 Specification K0463 Revision A EN English ...
Page 68: ...62 Specifiche K0463 Revisione A IT Italiano ...
Page 82: ...76 Especificações K0463 Revisão A PT Português ...
Page 96: ...90 Технические характеристики K0463 Редакция A RU Русский ...
Page 109: ......