• The cover can be cleaned by hand. To do this, use a 

moist (not wet) cloth and a mild cleaning solution.

• You can detach the cover using the zipper and wash 

it at a max. temperature of 60° using a household 
detergent.

• Do not dry the mattress cover in the dryer.
Note! If the Med Aire DuoWave is machine-washed 
and -disinfected, a validated and automated method 
according to the Robert Koch Institute (RKI) and/
or the Association for Applied Hygiene (VAH) is 
necessary.

Reuse/patient change

• In the case of reuse/patient change, the Med 

Aire DuoWave must be disinfected and cleaned 
according to a validated method recognized by the 
Robert Koch Institute.

• During processing, take note of the medical device 

law, the "Guidelines for the Processing of Medical 
Devices" as well as the "Guidelines for Hospital 
Hygiene and Infection Prevention" of the Robert 
Koch Institute.

Maintenance, service and storage

Maintenance
This product is subject to the Medical Device Law 
(MPG) and the Medical Devices Operator Ordinance 
(MPBetreibV). Internationally, the respective valid 
national regulations apply.
The functionality and proper condition must be checked 
before each use.
The following points must be regularly checked:
•  Check power cord and plug for wear and damage.

•  Check the housing of the unit for damage.

•  

Check the alternation between air supply and 
ventilation in each air chamber circuit. This alternation 
occurs approx. every 10 min. 

•  Check all hoses and connections for leaks, kinks and 

breakage.

• Check all LEDs which must light up in their respective 

function.

Please contact your specialist dealer if you identify any 
damage or defects.
During all visual inspections and maintenance, the unit 
must be disconnected from the power supply.
Testing and assessments must be performed and 
documented by suitable technical staff in accordance 
with DIN EN 62353.

Transport function

Remove the hose system from the unit and promptly 
close it with the attached cap. Now the mattress is in 
transport mode. To return to normal operation, remove 
the cap on the connection hose and connect it back to 
the unit and start the unit. 
Note: No patient should be positioned on the system 
when it is in transport mode!

Power failure

In the event of a power failure, set the system to transport 
mode. See section on transport function.

Alarm function

Low pressure
The "Low Pressure" LED is lit whenever the unit detects 
too little air in the cells. An acoustic signal sounds at the 
same time. Once the set pressure has been reached in 
the cells again, the display goes out and the acoustic 
signal stops. 
Important: During the filling phase, the "Low Pressure" 
LED is lit; once the pressure in the air chambers has 
reached the set user weight, the "Low Pressure" LED 
goes out.

Cleaning, disinfection and care

Unit
• Before cleaning, disconnect the unit from the power 

supply. During cleaning, ensure that no liquids/
cleaning solutions come into contact with the plug.

• The unit should not be saturated or soaked.
• The unit may be cleaned by hand. To do this, use a 

moist (not wet) cloth and a mild household cleaning 
and disinfectant solution.

• Do not use any solutions which contain phenol or 

abrasive cleaners. This can corrode the surface of the 
unit.

• Allow the unit to dry thoroughly.
• Perform a function test after the device is thoroughly 

dry.

Mattress overlay system
The mattress overlay system should not be saturated or 
soaked*, however it can be cleaned by hand. To do this, 
use a moist (not wet) cloth and a mild standard cleaning 
and disinfectant solution.
* There is a risk of liquid getting into the cell chambers 
and hoses.

6

Drive Medical GmbH & Co. KG | Subject to error and change

Instructions for use Med Aire DuoWave

Summary of Contents for DeVilbiss Med Aire DuoWave

Page 1: ...Instructions for use Med Aire DuoWave Alternating Pressure System ...

Page 2: ... Hand check 5 Alternating pressure function 5 Static function 5 Air flow 5 CPR cardiopulmonary resuscitation 5 Transport function 6 Power failure 6 Alarm function 6 Cleaning disinfection and care 6 Reuse patient change 6 Maintenance service and storage 6 Troubleshooting 7 Disposal 7 Warranty 7 Specifications 8 Type label 8 Service life 8 2 Drive Medical GmbH Co KG Subject to error and change Instr...

Page 3: ... Decubitus therapy up to and including grade 3 according to EPUAP Static pressure relief positioning is suitable for Patients who only occasionally tolerate dynamic alternating pressure or who do not tolerate it at all Contraindication Body weight less than 36 kg and over 150 kg In the case of unstable fractures particularly of the back cervical and lumbar spine In the case of neurological disease...

Page 4: ...on adjustable beds for disabled persons a sufficient distance between the top edge of the positioning system and the upper end of the bed rail must be ensured 220 mm according to DIN 1970 2000 If necessary corresponding side rail risers should must be attached to protect the patient Setting up the system TheMedAireDuoWaveisplacedontheexistingmattress Alternatively a 4 or 5 cm high foam pad can als...

Page 5: ...a 1 1 ratio An alternating cycle involves in each case the inflating and deflating of an air chamber and takes approx 10 minutes The first 3 air chambers in the head area remain static Static function The Med Aire DuoWave has a static function as a standard feature This is switched on using the Static button The LED indicates that the static function is enabled The air chambers all maintain the sa...

Page 6: ...nd assessments must be performed and documented by suitable technical staff in accordance with DIN EN 62353 Transport function Remove the hose system from the unit and promptly close it with the attached cap Now the mattress is in transport mode To return to normal operation remove the cap on the connection hose and connect it back to the unit and start the unit Note No patient should be positione...

Page 7: ...anty Damage due to material and or manufacturing errors will be corrected free of charge through a replacement or repair Troubleshooting Symptom Remedy Unit does not work Check whether the plug is correctly inserted Switch on the unit again If the LED on the On Off switch does not light up check the voltage of the outlet or insert the plug into another outlet If the LED on the On Off switch lights...

Page 8: ...e 2 Article number 3 Serial number 4 Production date year month 5 Voltage 6 DIN standard 7 Follow the instructions for use 8 CE marking 9 Type BF 10 Protection type 11 Protection class 1 and 2 12 Do not dispose of in household waste 13 Important Service life Our company assumes a long service life for this product as long as the product is used as intended and all maintenance and service requireme...

Reviews: