Drive DeVilbiss Healthcare Theia Instructions For Use Manual Download Page 13

13

7. INSTALLATION

• 

Ensure the support surface is used with a compatible side rail 

and bed frame combination – Incorrect combinations can lead to 

entrapment and/or falls hazards.

• 

Ensure the  support surface is of the correct type for the patient – 

Incorrect specification could lead to an injury. 

• 

The mains plug is the disconnect device for the means of isolating 

the control unit from the mains supply, the plug must be accessible 

at all times.

• 

Ensure the mains cable is plugged into an appropriate power source 

at all times.

• 

To avoid the risk of electric shock, this equipment must only be 

connected to a supply mains with a protective earth.

• 

Ensure the mains cable is not in tension, paying particular attention 

to when the bed travels up/down. 

• 

Precautions are to be taken when routing the mains cable around 

the bed to ensure that it does not become squeezed, trapped or 

damaged by the bedframe or other ancillary equipment - Risk of 

electrocution.

• 

Any electrical cable that is part of the system or associated 

ancillary equipment that is found to be damaged must be replaced 

immediately - Damaged electrical cables can create a risk of 

electrocution and/or fire.

• 

A CE marked extension cable must only be used when it is not 

possible to reach a wall socket with the equipment mains cable – 

Contact Drive DeVilbiss Healthcare Ltd. for detail in regards to safe 

use of extension cables. 

• 

If an extension cable is used never overload it by plugging in 

appliances that together will exceed the maximum current rating 

stated for the extension cable – Risk of fire. 

• 

Block adaptors are not to be used. 

• 

Ensure multiple socket outlets are not positioned under the bed 

frame - Liquids that leak onto such a socket could pose an electrical/

fire risk. 

• 

Consideration is to be taken in the positioning of the mains cable 

and air hose to minimise the risk of accidental strangulation resulting 

from patient, baby or child entanglement.

• 

Keep away from sources of heat and naked flames (e.g. cigarettes, 

fireplaces, electric fires, fan heaters, electric blankets etc.) – Close 

proximity could damage the electrical system and/or cover, bedding 

could catch fire etc.

• 

Do not place any objects or items, such as blankets, on or over the 

control unit - Risk of fire.

• 

Avoid placing the system in a moisture rich environment - Prolonged 

exposure to moisture could damage the electrical system and pose 

an electrical/fire risk. 

Warning

Summary of Contents for DeVilbiss Healthcare Theia

Page 1: ...Dynamic System Instructions for use...

Page 2: ...o normal use QUICK REFERENCE GUIDE P1 of 2 FIG 1 FIG 2 FIG 3 FIG 4 Connect the mattress cushion to the control unit FIG 5 ensuring the feed pipes do not kink or become trapped between bed chair parts...

Page 3: ...and cushion MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 No Symbol Description 3 MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 Static Mode Press to set...

Page 4: ...1 Preparing for Use 8 2 Control Interface 8 3 CPR Function 8 4 Low Pressure Warning 8 5 Power Failure Warning 9 DECONTAMINATION CLEANING 9 1 Control Unit 9 2 Mattress and Cushion 9 3 Alternative Cove...

Page 5: ...ilbiss Healthcare Ltd service engineers or qualified personnel 1 INTRODUCTION 2 CONTACT INFORMATION Drive DeVilbiss Healthcare Ltd Sidhil Business Park Holmfield Halifax HX2 9TN United Kingdom For Ser...

Page 6: ...as it depends on the physical size of the patient in relation to the proportions of the mattress and bed frame 3 3 Intended Use The intended use of the support surface is for sleeping resting sitting...

Page 7: ...tures 3 5 Indications To assist as part of an overall programme of care when active load distribution through mechanical means is required Bed frames used with the system can vary greatly depending on...

Page 8: ...dema when using alternating mode only Other contraindications may be relevant which are specific to the patient or care environment 4 4 Training If these instructions for use are not deemed sufficient...

Page 9: ...A hazard could be introduced Occupants and users of this equipment must never smoke in close proximity to the control unit mattress cushion or bedding being used with it Risk of fire Accessories that...

Page 10: ...could be incurred Do not store whilst inflated damage could be incurred Do not store objects such as side rails on top of the mattress or cushion damage could be incurred Warning Caution 5 2 Transport...

Page 11: ...e of potential product damage Manufacturer Date of manufacture Lot number Class II electrical device Theuser occupantisprotectedbyatleasttwolayersofinsulationbetweenthecurrent carrying parts e g contr...

Page 12: ...ts Type BF Applied parts which are electrically isolated from earth and other parts of the medical equipment Complying with specific requirements for protection against electric shock to EN 60601 1 95...

Page 13: ...electrical cables can create a risk of electrocution and or fire A CE marked extension cable must only be used when it is not possible to reach a wall socket with the equipment mains cable Contact Dr...

Page 14: ...imum of 2 hours prior to plugging into the mains supply Risk of electrical system damage if operated outside of the recommended temperatures Caution Open all packaging with care Once removed from the...

Page 15: ...control unit ensuring the air hose does not kink or become trapped between parts of the bed frame Plug the mains cable into a suitable mains supply and switch on the control unit The system will star...

Page 16: ...ay It s recommended to adjust comfort control to the max when inflating before adjusting the pressure to the patient s needs Before changing the comfort control a clinical judgement is required from f...

Page 17: ...connections are secure and correctly installed according to the relevant instructions If the pressure is consistently low open the zipper and confirm that all of the hoses are properly connected Chec...

Page 18: ...of the below washing instructions are not followed the product warranty will be invalidated Do not use solvents neat bleach phenolic based cleaning solutions or abrasive products to clean the casing...

Page 19: ...ter and a clean cloth and allow to fully dry before use Ensure the internal foam and air cells are orientated correctly to the mattress covers when reassembling 9 3 Alternative Cover Cleaning Instruct...

Page 20: ...ble and visual indicators work appropriately when plugged in and unplugged from mains supply Check that the battery is still functional and operates in the event of a power loss Check that the mattres...

Page 21: ...used with your system is not to be disposed of in general municipal waste as it is to be considered as Waste Electrical and Electronic Equipment WEEE Some of the electrical components could be harmfu...

Page 22: ...Cushion Weight 2kg Maximum Patient Weight 120kg 18 stone No of Cells 6 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 10...

Page 23: ...Mattress Weight 9 7kg Maximum Patient Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x...

Page 24: ...Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 100mm Control Unit Weight 1 6kg Cove...

Page 25: ...ed for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacture s declaration electromagnetic immunity Immunity...

Page 26: ...For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter...

Page 27: ...urer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in ea...

Page 28: ...ow instructions or g otherwise used in a manner inconsistent with any instructions provided by Drive DeVilbiss Healthcare Ltd The warranty explicitly exempts consumable items Thiswarrantycontainstheen...

Page 29: ...29...

Page 30: ...Airflo xiamen Medical Co Ltd 1F 3F 4F No 6 East Haijing Road Haicang Xiamen Fujian China Y Sung Handelsvertretung Duesselihaler Str 24 40211 Duesseldorf Germany Drive DeVilbiss Healthcare Ltd Sidhil...

Reviews: