SPECIFICATIONS
DEVILBISS 10-LITER SERIES
Catalog Number
1025DS
1025KS
1025UK
Delivery Rate
2 to 10 LPM
2 to 10 LPM
2 to 10 LPM
Maximum Recommended Flow
(@
nominal outlet pressures of zero and
7 kPa)**
10 LPM
10 LPM
10 LPM
Outlet Pressure
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Auxiliary Oxygen Port **
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Electrical Rating
120 V, 60 Hz, 6 .1 Amp
230 V~, 50 Hz, 3 .2 Amp
240 V~, 50 Hz, 3.2 Amp
Operating Voltage Range
102-132 V~, 60 Hz
195-253 V~, 50 Hz
204-264 V~, 50 Hz
Oxygen Percentage
2-10 LPM=87%-96%
2-10 LPM=93% +3%/-6%
2-10 LPM=93% +3%/-6%
Operating Altitude
(tested at 70˚F {21˚C} only)
0-1500 M (0-5000 ft)
Across the voltage range:
No degradation of performance
Across the voltage range:
No degradation of performance
Across the voltage range:
No degradation of performance
Operating Environment Range*
41˚F (5˚C) to 95˚F (35˚C), humidity
range of 15% to 93% non-
condensing
No degradation in performance across the
operating voltage range .
No degradation in performance across
the operating voltage range .
No degradation in performance across
the operating voltage range .
Power Consumption
120 vac, 60HZ: 639 watts average
230 vac, 50Hz: 664 watts average
240 vac, 50Hz: 670 watts average
Weight
42 lbs. (19 Kilograms)
42 lbs. (19 Kilograms)
42 lbs. (19 Kilograms)
Safe Working Load
53 lbs. (24 Kilograms)
53 lbs. (24 Kilograms)
53 lbs. (24 Kilograms)
Sound Pressure Level at 3 and
10LPM
<59 dBA
<57 dBA
<57 dBA
Sound Power Level at 3 and 10 LPM
<69 dBA
<67 dBA
<67 dBA
Dimensions
24.5”H x 13.5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
24.5”H x 13.5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
24.5”H x 13.5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
Maximum Limited Pressure
35 PSIG (241 kPa)
35 PSIG (241 kPa)
35 PSIG (241 kPa)
Operating System
Time Cycle / Pressure Swing
Time Cycle / Pressure Swing
Time Cycle / Pressure Swing
Low Oxygen Indicator
<82% low oxygen
<60% very low oxygen
<82% low oxygen
<60% very low oxygen
<82% low oxygen
<60% very low oxygen
Storage Conditions
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Equipment Class and Type
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IP21
Class II Equipment Double Insulated;
Type B Applied Part
IP21
Class II Equipment Double Insulated;
Type B Applied Part
IP21
Approval Body and Safety Standard
TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012)
+A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014
CAN/CSA-C22.2 No. 60601-1:14
CAN/CSA-C22 .2 No . 60601-1-6:11
CAN/CSA-C22 .2 No . 60601-1-11:15
CAN/CSA-C22 .2 No . 80601-2-69:16
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
CE mark
No
Yes
Yes
EMC Compliance To
EN60601-1-2
EN60601-1-2
EN60601-1-2
* NOTE
– The OSD performance at 41˚F (5˚C) to 95˚F (35˚C), 93% R.H. through voltage range on the 1025 verified at 670m.
**The maximum recommended flow is 6 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.