HPG 13150 400 V (A)
Communication Interfaces
4-53
HPG™ 13150 400 V RF Power Generator
Prerelease version
96
set pulsing
duty cycle
Sets the RF pulsing duty ON time in
increments of 1%. This value can range
from 1% to 99%.
• Minimum On or Off time is 16 µs.
See also commands
27
and
93
. For more
information, see
Read back with command
196
.
2
1
(CSR only)
111
initialize
capacitors
Moves capacitors of a connected
Variomatch unit to the minimum position.
(For more information on this interface,
see
“Matching Interface” on page 4-2
.)
This command will not work with other
matching networks unless they are
electrically compatible and are connected
through the
Matching Interface
connector.
0
1
(CSR only)
112
move load
capacitor
position
Moves the load capacitor motor of a
connected Variomatch unit to the
specified position. (For more information
on this interface, see
.) This command
will not work with other matching
networks unless they are electrically
compatible and are connected through
the
Matching Interface
connector.
Send two data bytes to move the load
motor to its new position.
• Accepts a value of 0 to 1000
Read back with command
175
.
2
1
(CSR only)
Table 4-13. Host port commands (Continued)
Command
Description
Number of
Host Data
Bytes
Number of
Response
Data Bytes
Summary of Contents for 13150 400 V
Page 2: ......
Page 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Page 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...