background image

 

 

- 12 -

 

A. Press  “▲” and  “▼”  to select  the  previous 

“20”

  programs. (The  selected program  will  be operated; if  no 

selection is made, the previously operated program will be operated.) 

B. Place the  sample (boiling  tube).  If no  boiling tube  is placed, the  program  will not be performed  to avoid 

reagent or steam overflow.   

C. Press 

“ENT”

 to operate the automatic distillation program, and the apparatus will run in sequence of “water 

filling——alkali filling——acid filling——distillation (countdown)——ending”. 

D.  After  the  set  distillation  time  (the  distilled  solution  is  about  100mL,  with  the  duration  of  about  400S)  is 

ended, the “ticking-” alarm sound with about 2S will be sent, and the program is completed. 

E. During the distillation process, 

“ALM” 

can be pressed to exit from the distillation program.   

Note:  Due  to  different  cooling  water  temperature  and  flow  of  the  places  (users),  the  time  for  distilling  the 

liquid  with  same  volume  (such  as  100ml)  may  be  different;  the  users  are  recommended  to  perform  trial 

distillation for several times according to the actual conditions to master (set) the distillation duration. 

 

 

VII. Reagent Preparation Method 

7.1 Prepare 2L 30%~40% sodium hydroxide solution and pour it into the alkali barrel (suggestion: The 35% 

sodium hydroxide solution is recommended, for it will not easily crystallize and block the pipeline in case of 

changeable room temperatures). 

7.2 The methyl red -bromocresol green mixed indicator (preparing based on the sector standard). 

7.3 Prepare 2L 2% boric acid solution and pour it into the acid barrel; the mixed indicator and boric acid can 

be mixed based on the sector standard and poured into the acid barrel. The mixed indicator can be mixed after 

distillation and before titration. 

Note: If the titration is applied with the ZDG-1automatic luminosity titrator; only the methyl red is required 

for the indicator. 

 

 

VIII. Precautions for Use of Apparatus 

8.1  The  preheating  is  needed  after  starting  the  apparatus;  after  that,  directly  enter  the  manual  operation  and 

Summary of Contents for DW-ATN-300

Page 1: ...N 300 Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Page 2: ...s Structure 6 V Installation Method 7 5 1 Pre installation Check 7 5 2 Installation Conditions 7 5 3 Installation Steps 7 5 4 Debugging 7 VI Application Methods and Steps 8 6 3 Flow Calibration 9 6 4 Program Settings 9 6 5 Manual Distillation 10 6 6 Automatic Distillation 11 VII Reagent Preparation Method 12 VIII Precautions for Use of Apparatus 13 ...

Page 3: ...ctions Danger Property or personal injuries may be caused please use the machine as required Special power supply Reliable grounding Flammable and explosive No transformation 1 Please use the power supply and voltage conforming to the product requirements 2 The applicable voltage for the equipment is AC220V 50HZ 3 The product shall be reliably grounded and kept away from electromagnetic interferen...

Page 4: ...7 The working environment shall be free from dust and flammable and explosive gases 8 The working environment shall be of 10 45 and relative humidity less than or equal to 80 9 The apparatus shall be installed in the place without direct sunlight 10 As the reagent and other solution are corrosive safety appliances shall be equipped during operation for safety ...

Page 5: ...y Card 1 Product Qualification Certificate 1 Accessories in the accessory box Distilled water inlet pipe 6mm 1 5m 1 Alkali liquid inlet pipe 6mm 1 5m 1 Boric acid solution inlet pipe 6mm 1 5m 1 Waste distilled water pipe 6mm 1 5m 1 Condensed water inlet and outlet pipes 8mm 2 5m 1 for each Digestion pipe 40 300mm 1 Gloves 1pair Power line 1 8m 1 ...

Page 6: ... combinations for water filling alkali filling acid filling and distillation time for setting and application by the user ATN 300 has applied the British self sucking pump method for liquid feeding rather than the inflatable method applied by many domestic manufacturers which can place the liquid feeding barrel under the control console and select the liquid feeding barrel with large capacity ATN ...

Page 7: ...water consumption 1 2L 2 4L min 3 5 Amounts of water filling alkali filling acid filling 0 99ml 3 6 Sample weight Solid 6g liquid 30mL 3 7 Operation mode Dual modes of manual distillation automatic distillation 3 8 Automatic distillation mode Ten kinds of distillation programs 3 9 Alarm and shutdown temperature 120 3 10 Power voltage AC220V 50HZ 3 11 Maximum power 1600w 3 12 Maximum current 7 5A 3...

Page 8: ...wer switch 6 Distillate outlet pipe boric acid filling pipe 7 Conical flask 8 Digestion pipe tray 9 Acid liquid barrel 10 Alkali liquid barrel 11 Distilled water barrel Figure of the Back side Structure of the Apparatus see Figure 4 2 1 Distilled water inlet 2 Alkali liquid inlet 3 Boric acid solution inlet 4 Waste liquid outlet 5 Condensed water outlet 6 Condensed water inlet 7 Power socket The m...

Page 9: ...8 C 6 The drain tank shall be 10cm lower than the apparatus outlet to ensure a smooth drainage 7 As the equipment is a kind of precise apparatus acute vibration or transverse and upside down placement shall be avoided 5 3 Installation Steps refer to Figure 4 1 and Figure 4 2 1 Place the apparatus on a stable experiment table 2 Connect one end of the condensed water inlet pipe with the condensed wa...

Page 10: ...rtup is shown in the Figure 6 1 Function selection 6 2 Press key to enter the function selection page See Figure 6 2 Figure 6 2 Function selection F1 Correction Flow correction F2 Manual Manual distillation F3 Automatic Automatic distillation F4 Program Program setting 6 3 Flow Calibration Prior to flow calibration connect with the container after that press F6 F7 F8 keys for several times to fill...

Page 11: ...the operation is ended record the distilled water volume in the graduated cylinder C Boric acid flow calibration Place the distillate outlet pipe into the graduated cylinder press F7 key to select the acid flow calibration After the operation is ended record the boric acid solution in the graduated cylinder D Press SET after above steps are completed E Input the recorded volume mL value into the c...

Page 12: ...li 12ml Alkali filling amount 12 ml Acid 12ml Acid filling amount 12 ml Time 123s Distillation time 123 S A Press SET key to perform program parameter settings B After each setting press ENT once for confirmation using the following sequence C SET Program X selection ENT Water filling amount ml ENT Alkali filling amount ml ENT Acid filling amount ml ENT Distillation time S ENT ALM Remember After p...

Page 13: ...rement press F3 again to stop the distillation D Press F4 to select the water filling when the solution in the digestion pipe reaches the requirement press F4 again to stop the water filling E Press F5 drainage means to drain out water in the steam generator When impurities occur in the steam generator or the apparatus is not used for a long time water in the steam generator shall be changed or dr...

Page 14: ...olume such as 100ml may be different the users are recommended to perform trial distillation for several times according to the actual conditions to master set the distillation duration VII Reagent Preparation Method 7 1 Prepare 2L 30 40 sodium hydroxide solution and pour it into the alkali barrel suggestion The 35 sodium hydroxide solution is recommended for it will not easily crystallize and blo...

Page 15: ...k the rubber plug 8 4 As the digestion pipe will be very hot after distillation gloves shall be worn upon taking it down to avoid scalding 8 5 The condensed water must be opened during distillation 8 6 The filled alkali must ensure the alkalinity of the digestion solution greater than or equaling to 11 8 7 If the apparatus is not used for a long time in winter water in the steam generator must be ...

Page 16: ...rnational Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Reviews: