recommended to perform re-calibration according to actual conditions.)
6.4. Program Settings
The “program” refers to different “combinations” for “water filling, alkali filling, acid filling, and
distillation time”, and the machine can save20 kinds of different “combinations”.
Press “F4” in the “function selection” page to enter the “program settings” page: See Figure 6-5.
Figure 6-5
Program Settings
Water 12 ml: Water filling amount 12 ml Alkali 12ml:Alkali filling amount 12 ml
Acid 12ml:Acid filling amount 12 ml Time 123s:Distillation time 123 S
A. Press
“SET”
key to perform program parameter settings.
B. After each setting, press
“ENT”
once for confirmation, using the following sequence:
C.SET ▲, ▼Program X selection ENT Water filling amount ml ENT
Alkali filling amount ml ENT Acid filling amount ml ENT
Distillation time S ENT
ALM
.
(Remember!!! After pressing “ENT” after completing each program setting, “ALM” must be pressed for
saving; or else, the program is invalid.)
6.5. Manual Distillation
Press “F2” in the “function selection” page to enter the “Manual Distillation” page: See Figure 6-6.