3-4
Checking Input Range
Checking Input Range
Input range
The Input Range screen is displayed by pressing the CkRng key on the Scope mode
screen. Input Range shows the detected range for all channels. The number after the
status shows the percentage of the input of the instrument’s full range.
The input range can only be viewed, not changed. Press OK when done viewing the
input range.
Input range
description
The following table describes the messages that may register on the Input Range
screen. The instrument is able to detect the input range for voltage and current channels
A, B, C, and D.
If input range is OVER, UNDER, or CLIP, then the rms readings will not be accurate.
Using CTs below 10% of Full Scale may also cause inaccuracies in amplitude, phase
and harmonic distortion readings.
Input Range
Description
OK
Within Range
OVER
Over Range
UNDER
Under Range
N/A
Channel Disabled
CLIP
Clipping
MARK205
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Summary of Contents for PowerXplorer PX5
Page 68: ...3 26 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 84: ...4 16 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 166: ...7 22 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 212: ...A 10 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 226: ...B 14 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 232: ...C 6 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 234: ...D 2 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 258: ...F 4 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...