Dräger Zeus CLIC Abridged Instructions Download Page 2

2

Abridged Instructions for 
Zeus CLIC Adapter

These Abridged Instructions are a supplement to the 
Instructions for Use for the Zeus SW 3.n Anaesthesia 
Workstation System

Follow the Instructions for Use for the Zeus SW 3.n 
Anaesthesia Workstation System.

Follow the Instructions for Use for the disposable absorber 
(Drägersorb CLIC 800 Plus or Drägersorb CLIC Free).

Intended Use

The Zeus CLIC Adapter facilitates use of the disposable 
absorber Drägersorb CLIC 800 Plus or 
Drägersorb CLIC Free.

What‘s what

1

Housing for the IBF Absorber Filter

2

Sight glass, orange display when an absorber filter 

is installed

3

 Release button with bypass function

4

Support for the disposable absorber

Preparation

When present, turn the absorber anticlockwise and 
remove downwards. 

Fit the Zeus CLIC Adapter

Insert the new IBF Absorber Filter in the Zeus CLIC Adapter 
from above using the recessed handles.

Attach the Zeus CLIC Adapter to the breathing system 
from below and turn it clockwise as far as it will go. 

Make sure that the release button and the sight glass are 
facing forwards and visible to the user. 

Carry out the leak test and self test with the disposable 
absorber engaged – see Zeus Instructions for Use. 

Changing disposable absorbers during operation

Proceed according to the User Instructions of the 
disposable absorber being used.

Caution!

When operating the unit without the CO

2

 monitor 

(instrument switched off or sample line blocked) adjust the 

fresh-gas flow larger minute volume so as to prevent the 

inspiratory CO

2

 concentration from increasing. 

Caution!

No CO

2

 is absorbed during the changeover. Monitor the 

inspiratory CO

2

 and, if necessary, adjust the fresh-gas flow 

to greater than the minute volume so as to prevent the 

inspiratory CO

2

 concentration from increasing.

00

1

1

2

3

4

Service

Change the IBF Absorber Filter after 20 operations or after 
20 hours, whichever comes first.

Disassembling the adapter

The adapter is disassembled by following the steps for 
assembling it in reverse order.

Caution!

After changing the absorber, check to make sure that the 
breathing system is positioned horizontally on the support 
in order to prevent any loss of gas.

Caution!

If used without an IBF Absorber Filter, contamination of the 

turbine cannot be ruled out.

en

Summary of Contents for Zeus CLIC

Page 1: ...triegelungstaste und Schauglas nach vorn sichtbar zum Anwender zeigen Leckage und Selbsttest mit eingeklinktem Einwegabsorber durchf hren siehe Gebrauchsanweisung Zeus Einwegabsorber w hrend des Betri...

Page 2: ...ton and the sight glass are facing forwards and visible to the user Carry out the leak test and self test with the disposable absorber engaged see Zeus Instructions for Use Changing disposable absorbe...

Page 3: ...de d gagement et le voyant se trouvent l avant et sont visibles par l utilisateur Effectuer le test de fuite et l auto test alors que l absorbeur jetable est engag se reporter la notice d utilisation...

Page 4: ...lante y est n visibles al usuario Realice la prueba de fugas y el autochequeo con el absorbedor de un solo uso encajado consulte las Instrucciones de uso Zeus Cambio de absorbedores de un solo uso dur...

Page 5: ...dat de ontgrendelknop en het kijkglas naar voren wijzen en zichtbaar zijn voor de gebruiker De lektest en zelftest uitvoeren terwijl de disposable absorber goed op zijn plaats zit zie de gebruiksaanw...

Page 6: ...siano rivolti in avanti e siano visibili dall utente Eseguire il test perdite e l autotest con l assorbitore monouso innestato Vedere le istruzioni per l uso di Zeus Sostituzione degli assorbitori mo...

Page 7: ...tglaset r riktade rakt fram och synliga f r anv ndaren Utf r l ckagetest och sj lvtest med absorbern f r eng ngsbruk aktiverad se bruksanvisningen till Zeus Byte av absorber f r eng ngsbruk under drif...

Page 8: ...o visor est o virados para cima e vis veis para o utilizador Efectue o teste de exist ncia de fugas e o auto teste com o absorvedor descart vel encaixado consulte a Instru es de Utiliza o do Zeus Troc...

Page 9: ...ellen rz ablak el re n z e s a felhaszn l sz m ra l that e Az egyszer haszn latos abszorber felszerel se ut n v gezzen sziv rg s ellen rz st s futtassa le az nellen rz st Ennek menet t l sd a Zeus al...

Page 10: ...hybu hodinov ch ru i ek Ujist te se e uvol ovac tla tko a pr zor sm uj dop edu a jsou pro u ivatele viditeln Prove te test t snosti a autotest s osazen m jednor zov m absorb rem viz N vod k pou it p s...

Page 11: ......

Page 12: ...ical products Dr ger Medical GmbH Germany Moislinger Allee 53 55 D 23542 L beck 49 451 8 82 0 FAX 49 451 8 82 20 80 http www draeger medical com 90 38 856 me Dr ger Medical GmbH Edition 3 2010 09 Subj...

Reviews: