Wiring diagram for 
connecting non-intrinsically 
safe quipment to AAC 00xx 
or AAC 03** 

 

 

Notes 

 

 

1. XEXT1 and/or XEXT2 can be used 

for connection of non-intrinsically 
safe equipment via Z-barriers. As 
an example the wiring diagram 
shows a connetion to XEXT2. 
 

2.  Series switching of the types 

Fuchs Z757

 with 

Pep-

perl+Fuchs Z966H

 (BAS 01 ATEX 

7005 and IECEx S06.0012 xx) is 
permitted as shown in the following 
diagrams. 
 

3.  Barrier must be installed as in-

structted by the manu

facturer’s 

control drawing. 
 

4. 

Caution

: Use connection XEXT1 

either for charging (pin 4,5) outside 
the hazardous area or the RS485 
port (pin 1, 2, 3). Using XEXT1 for 
both applications at the same time 
is not permitted. 
 

5. 

Caution

: Use connection XEXT2 

either for the relay output (pin 4, 5, 
6) or the RS485 port (pin 1, 2, 3).  
Using XEXT2 for both applications 
at the same time is not permitted. 
 

6. 

Warning: 

Substitution of compo-

nents may impair intrinsic safety. 
 

7. 

Cable Parameters: 

Use Belden 

3107A, 2 x 2 AWG22. Maximum 
length = 25 m = 82 ft between 
AAC 00xx and the Z-barriers. 

Title

Document Number - Revision

Page

&6$&RQWURO'UDZLQJ$$&[[$$&

6(  

 

$33529$/

S

c

hutz

v

er

me

rk IS

O 1

6

0

16 

be

ach

ten

.

Use o

f th

is

 d

o

cu

me

nt /

 co

nte

n

ts is

 fo

rbi

dde

n

with

o

u

e

x

p

ress

ed writt

en au

th

o

rity.

 All righ

ts

 r

e

se

rve

d

.

Dräg

er

„Ex i“ –

Power Supply 

(4 infallibly segregated circuits)

Uo-1, Uo-2, Uo-3, Uo-4 <= 13.10 V

Io (spark)-1, Io (spark)-2, Io (spark)-3, Io (spark)-4 <= 0.850 A

I (therm)-1, I (therm)-2, I (therm)-3, I (therm)-4 <= 0.425 A

Po-1, Po-2, Po-3, Po-4 <= 2.78 W

Co-1, Co-2, Co-3, Co-4 <= 1 nF

Lo-1, Lo-2, Lo-3, Lo-4, <= 1 uH

Notes

1.

Caution:

Only connect galvanically isolated, intrinsically safe electric 

circuits of protection type Ex ia IIC with linear output characteristics 
that meet the safety parameters of AAC 03**.
All of the 4 circuits must be infallibly isolated from each other.

2.

Caution:

Operation with an intrinsically safe power supply to XEXT3 

is only permitted without simultaneous inductive charg-ing, or 
simultaneous power charging via the XEXT1 connection.

3.

Warning:

Live maintainance is not permitted in Zone 0 areas.

4.

Warning:

Substitution of components may impair intrinsic safety.

5.

Cable Parameters

If cable parameters are unknown, use the following parameters:

Cable Capacitance:  C

c

= Ccable = 60 pF/ft (200 pF/m)

Cable inductance: L

c

= Lcable = 200 nH/ft  (660 nH/m)

The cable must maintain infallible segregation of the 4 circuits.

Supply of AAC 03** in Hazardous Areas

XEXT3

for 

each 

circuit

Ex i Power 
Supply

Uimax

13.10 V

>=

Uo

Iimax

0.850 A

>=

Io

Pimax

2.78 W

>=

Po

Ci

1 nF

+C

c

<=

Co

Li

1 uH

+L

<=

Lo

Hazardous Area

Safe Area

4 x 2

Use only connector CNLINKO 
LP-20-J09PE-01-22 or equivalent
Manufacturer:
Shenzen Linko Electric Co. Ltd.
3rd Floor, Building A,
Pengzhou Industrial Park
Fuyuan 1st Road, Fuyong Town,
Baoan District SZ.China