Troubleshooting
Dräger Quaestor 5000/7000
21
4
Troubleshooting
5
Maintenance
5.1
Maintenance intervals
5.2
Cleaning
If required, clean the test device with a moist leather cloth
and dry it.
5.3
Maintenance work
5.3.1
Replacing the gel face
1. Remove the old gel face from the test head.
2. Moisten the inside of the new gel face with water.
3. To place the new gel face onto the test head, first position
the eye with the measuring point on the test head, and then
position the other eye. Press the gel face against the
test head.
4. Ensure that the edge of the gel face is flush around the
entire test head.
5. Dräger recommends mounting a full face mask to the test
head until the moisture between the test head and the gel face
has evaporated (e. g. over night) and then commencing
the test.
6
Transport
Only transport the test device when it is switched off. The test
device may not be used during transport in a vehicle.
7
Storage
Store the test device in a dry place. Avoid direct sunlight and
temperatures above 60 °C.
8
Disposal
Fault
Cause
Remedy
The test device
makes a hissing
sound when it is
switched on.
Test device faulty
Contact service
personnel
The LED on the
test device does
not light up, even
though the test
device is
connected to
power supply and
a test has been
opened.
Test device faulty
Contact service
personnel
The LED on the
power supply unit
does not light up.
Power pack faulty
Contact service
personnel
No communication
with the PC
software is possible
(the test device is
not recognised).
USB cable faulty
Use a new cable
Test device faulty
Contact service
personnel
Work to do
Af
te
r use
Annua
lly
Ev
ery
6
ye
ars
Test the gel face
X
Conduct inspection
1
1
Only by trained specialist personnel.
X
Conduct a major overhaul
2
2
To ensure the availability of test devices and thus support the
workflow in respiratory protection workshops, major overhauls on
test devices should be carried out as required by Dräger. Based on
the various uses for the test devices, test devices are subject to
different types of wear, but essentially Dräger recommends having
a major overhaul conducted at least every six years.
X
This product is not permitted to be disposed of as
domestic waste. It is therefore marked with the
adjacent icon.
Dräger will accept the return of this product free
of charge. For information contact the national sales
organisations and Dräger.