background image

3362443 (A3-D-P) – Page 1 of 4

PSS 5000 Series

Self-contained breathing apparatus with Sentinel TX Gauge

Instructions for Use

i

N

N

1

4552

2

3

5

6

1

5

3

6

4

4344

3379

3088

4493

1

7

3094

1

8

3095

2

3

D

D

For your safety

1.1  General safety statements

 

● Before using this product, carefully read the Instructions for Use.

 

● Strictly follow the Instructions for Use. The user must fully understand 

and strictly observe the instructions. Use the product only for the 

purposes specified in the Intended Use section of this document.

 

● Do not dispose of the Instructions for Use. Ensure that they are 

retained and appropriately used by the product user.

 

● Only fully trained and competent users are permitted to use this product.

 

● Comply with all local and national rules and regulations associated 

with this product.

 

● Only trained and competent personnel are permitted to inspect, repair 

and service the product as detailed in these Instructions for Use (see 

Section 5). Further maintenance work that is not detailed in these 

Instructions for Use must only be carried out by Dräger or personnel 

qualified by Dräger. Dräger recommend a Dräger service contract for 

all maintenance activities.

 

● Use only genuine Dräger spare parts and accessories, or the proper 

functioning of the product may be impaired.

 

● Do not use a faulty or incomplete product, and do not modify the 

product.

 

● Notify Dräger in the event of any component fault or failure.

 

● Do not use any form of chemical marking or paint on the equipment.

 

● Use of the apparatus should be consistent with NFPA 1500 – Standard 

on Fire Department Occupational Safety and Health Program. 

 

● All  approved  respiratory  equipment  shall  be  selected,  fitted,  used, 

and maintained in accordance with MSHA (Mine Safety and 

Health Administration), OSHA (Occupational Safety and Health 

Administration), and other applicable regulations.

 

● The air supply shall meet the requirements for breathing air according 

to CGA G – 7.1, Grade D or higher quality and, where appropriate, be 

in accordance with: NFPA 1989 Standard on Breathing Air Quality for 

Emergency Services Respiratory Protection.

 

● This device has been tested and complies with the FCC (Federal 

Communications Commission) and IC (Industrial Commission) Rules. 

Changes or modifications not expressly approved by the manufacturer 

will void your authority to operate this product.

 

● Before occupational use of this respirator a written respiratory protection 

program must be implemented meeting all the local government 

requirements. In the United States employers must comply with 

OSHA 29 CFR 1910.134 which includes medical evaluation, training, 

and fit testing.

1.2  Definitions of alert icons

Alert  icons  are  used  in  this  document  to  provide  and  highlight  text  that 

requires a greater awareness by the user. A definition of the meaning of 

each icon is as follows:

!

WARNING

Indicates a potentially hazardous situation which, if not avoided, 

could result in death or serious injury.

!

CAUTION

Indicates a potentially hazardous situation which, if not avoided, 

could result in physical injury or damage to the product or 

environment. It may also be used to alert against unsafe practices.

i

i

NOTICE

Indicates additional information on how to use the product.

Description

2.1  Product overview

The Dräger PSS 5000 Series is a compact and lightweight breathing 

apparatus that provides the wearer with respiratory protection using an 

open-circuit, pressure-demand, compressed-air system. The apparatus 

can be used as a self-contained system, or with an independent air supply 

for supplied-air respirator (SAR) operations. The series is compatible 

with a wide range of Dräger compressed-air cylinders and face masks 

(e.g. FPS 7000 face masks, and composite cylinders). The product 

includes a Dräger Sentinel TX Gauge.

2.2  Feature description

The carrying system has a carbon-composite backplate, with adjustable 

shoulder straps and waist belt connected using quick release connectors. 

The waist pad is connected at a flexible joint to compensate for the twisting 

and bending of the user.

All variants use the same high-performance first-stage regulator (Fig 1, 

Item 5). The regulator supplies medium-pressure breathing air through a 

medium-pressure hose (6) and a quick coupling (2) to an attached lung 

demand regulator. Incorporated in the first-stage regulator is a RIC UAC 

(rapid intervention crew universal air connection) (4), which is a male 

coupling that allows emergency refilling of the air cylinder while the wearer 

is breathing from the apparatus.

The TX Gauge (Fig 2) is a mechanical pressure gauge (Bourdon tube 

type) that has a photoluminescent gauge face with LED illumination, and 

a whistle that sounds at the end-of-service-time (see Section 8 for the 

end-of-service-time indicator (EOSTI) activation pressures). In addition to 

the normal display of cylinder pressure, the TX Gauge provides battery 

powered visual signals that indicate that the gauge is switched on (green 

LED), low battery and hardware error alerts (amber LED), and low cylinder 

pressure (red LED).

2.2.1  TX Gauge radio frequency link

The TX Gauge is equipped with radio frequency (RF) communication 

circuitry that allows:

 

● Reading and reprogramming of the gauge

 

● Wireless transmission of data to compatible devices

 

● Pairing of the gauge with other compatible devices

Information that can be read from the TX Gauge includes device identity, a 

record of events (datalog), the current values for configurable parameters 

and the firmware version. Configurable parameters include intermediate 

pressure warning activation pressures, gauge illumination duration, etc. 

The parameters described in this document are the default for the device. 

Reading and reprogramming of the gauge requires the Dräger PC Link 

Module. Contact Dräger for details.

The TX Gauge has a short range RF wireless transmitter. This transmitter 

can transmit data (such as switch on/off signals, pressure values, etc.) to 

compatible devices (e.g. the FPS 7000 HUD (head-up display)) that are 

within communication range. See the device Instructions for Use or contact 

Dräger for details.

The TX Gauge may be paired with compatible removable integrated 

Dräger Personal Alert Safety System (PASS) devices. Pairing ensures that 

the PASS will only respond to signals transmitted by the paired TX Gauge, 

and will ignore switch on/off signals transmitted from any other devices in 

range. More than one PASS device may be paired with a single TX Gauge. 

To check for paired devices within range, press and hold the gauge 

illumination button for 10 seconds. See the PASS device Instructions for 

Use or contact Dräger for details.

2.2.2  Compressed-air cylinders, lung demand regulators and face 

masks

The PSS 5000 Series is compatible with composite material cylinders 

of 30 to 60 minute capacity, and is available in 2216 psi or 4500 psi 

versions. Full descriptions and user instructions are contained in separate 

instructions supplied with the cylinder, face mask or lung demand regulator.

2.3  Intended use

When the PSS 5000 Series breathing apparatus is used with an approved 

lung demand regulator, face mask and air cylinder, it provides a wearer 

with respiratory protection for working in contaminated or oxygen-deficient 

conditions.

The cylinder, lung demand regulator, face mask and other accessories 

used with this product must be certified Dräger components, assembled in 

an  approved  configuration,  otherwise  the  operation  of  the  device  may  be 

impaired. Contact Dräger for further information.

2.4  Approvals

The Dräger PSS 5000 Series is certified to 42 CFR Part 84 by NIOSH 

and in certain combinations to provide respiratory protection from military 

grade chemical, biological, radiological, and nuclear hazards (CBRN). The 

series is also certified by SEI to meet the requirements of NFPA 1981:2013. 

Approved PASS devices meet NFPA 1982:2013. The apparatus must only 

be used in conjunction with compressed-air cylinders approved by NIOSH.

The Dräger Sentinel TX Gauge is approved for use with Dräger NIOSH and 

NFPA 1981:2013 breathing apparatus, and has the following approvals:

 

● UL913 Class I, II, Div.1 Groups C-G

 

● Exia CAN/CSA C22.2 No.157

 

● T4 (Ta = -30 to +60 °C)

 

● This device (model: Sentinel TX Gauge) complies with part 15 of the 

FCC Rules. Operation is subject to the following two conditions: 1) This 

device may not cause harmful interference, and 2) This device must 

accept any interference received, including interference that may 

cause undesired operation.

 

● This device complies with RSS-310 of Industry Canada. Operation 

is subject to the condition that this device does not cause harmful 

interference.

2.5  Explanation of marking and symbols

!

CAUTION

Do not use marker pens or apply paint, and do not scratch or 

engrave the gauge, as this may damage the equipment.

Refer to the relevant authority for explanation of approval body symbols 

and marking on the equipment. Examples of other marking on component 

parts of the breathing apparatus are:

BRAC-1359 

– 

Dräger serial number

08/09 

– 

Month and year of 

manufacture

3356812 or R21034 

– 

Dräger part number

SF 

– 

Standard force coupling

LF 

– 

Low force coupling

 

Gauge illumination button

 

Green LED

 

Amber LED

 

Radio frequency (RF) 

communication symbol

Use

!

WARNING

Only trained and competent personnel should prepare and use 

breathing apparatus. Ensure that any accessories, ancillary equipment 

and other protective clothing items do not interfere with the apparatus 

and do not create a safety hazard.

The effective working duration of the apparatus is dependent on 

the initial air supply available and the breathing rate of the wearer. 

Fill air cylinders to their full rated pressure prior to use, and do not 

commence any operation (including supplied-air respirator (SAR) 

operations) using a cylinder that is less than 90 percent full.

!

CAUTION

Do  not  apply  excessive  force  or  use  tools  to  open  or  close 

a cylinder valve, and do not drop or throw down the breathing 

apparatus.

Refer to the following additional information before preparing or 

using the breathing apparatus:

 

● The special instructions (see Section 9).

 

● For non-CBRN use see the separate NIOSH Approval 

Label 3361125 for approved configurations.

 

● For CBRN use see the separate NIOSH CBRN Approval 

Label  3361124  for  approved  configurations.  For  CBRN  use 

the user must also refer to the FPS NFPA Lung demand 

Regulator and CBRN Special Instructions.

3.1  Preparation for use

3.1.1  Initial assembly (only required on receipt of the equipment)

The  battery  cover  and  batteries  are  supplied  loose  (unfitted)  with  the 

TX Gauge. Install them before first use (see Section 3.5.3).

9

4094

D

1

3

4

4093

2

1  Gauge illumination button

2  Active LED

3  Low battery LED

4  RF communication point

5  Photoluminescent gauge 

face

5

2

D

4

D

Summary of Contents for PSS 5000 Series

Page 1: ...ng of the gauge with other compatible devices Information that can be read from the TX Gauge includes device identity a record of events datalog the current values for configurable parameters and the firmware version Configurable parameters include intermediate pressure warning activation pressures gauge illumination duration etc The parameters described in this document are the default for the de...

Page 2: ...this may damage the equipment 1 Loosen the face mask straps As the seal between the face mask and the face is broken press the reset button Fig 3 Item 1 to switch off the positive pressure Fully remove the face mask and extend all of the straps of the head harness 2 Close the cylinder valve 3 Press the front button Fig 3 Item 2 to vent system and then press the reset button Fig 3 Item 1 to switch ...

Page 3: ...m the drying facility immediately when dry Drying time in a heated dryer must not exceed 30 minutes Do not immerse pneumatic or electronic components in cleaning solutions or water If water is trapped and then freezes inside the pneumatic system of the breathing apparatus such as the lung demand regulator its operation will be impaired Prevent any liquid from entering and thoroughly dry the breath...

Page 4: ...e and between multiple entries during the same use Decontamination and disposal procedures must be followed If contaminated with liquid chemical warfare agents dispose of the SCBA after decontamination U The respirator should not be used beyond 6 hours after initial exposure to chemical warfare agents to avoid possibility of agent permeation EBSS EBSS Activation or engagement of EBSS in either the...

Reviews: