background image

Intended Use

Intended Use

The Dräger Polytron Pulsar 2 Open Path Gas Detector is intended for stationary, 
continuous monitoring of hydrocarbon gases or vapours that may present and 
explosion hazard.

Explosion protection approvals

The explosion-protection approvals are valid for use of the device in gas/vapour-
air mixture of combustible gases and vapours under atmospheric conditions. The 
explosion-protection approvals are not valid for use in oxygen enriched atmospheres. 
In case of unauthorized opening of the enclosure, the explosion-protection approval 
is voided.

NOTE:  The Dräger Polytron Pulsar 2 Open path Gas Detector is one of the 

family of detectors certified under the designation GD8.  All certificates will 
refer to the GD8.

Parts Supplied

1. 

The Dräger Polytron Pulsar 2 Open Path Gas Detector comes in two parts, 
a transmitter and a receiver; each supplied with an ABS moulded cover.  
Additionally the detector is supplied with either:

— Flying leads and Ex e junction boxes.

— Ex d certified plugs and sockets.

2. 

An optical attenuator for use on beam paths between 4 and 16 meters, NOTE 
this is only supplied with 4-60m transmitters.

3. 

An installation sheet and a quick start guide are supplied with the detector. 
Installation/operation manuals are also available through local distributors or 
by contacting Dräger UK Ltd.

4. 

You will also require a commissioning kit which is ordered separately and 
includes:

— Hand Held terminal (HHT)
— 4mm Allen key
— Test Sheets
— HHT PC software
— PC Cable

Installing

Only trained service personnel (e.g. Dräger service personnel) may install the gas 
detector under observation of the regulations applicable at the respective location.

Choosing the path of the beam and Mechanical Installation

1. 

The siting of an open path gas detector is often not as critical as a point 

detector, since the released gas only has to find its way into some portion of 

the beam instead of to a particular point. However, siting is still an important 
consideration. Guidance for siting is contained in EN 60079-29-2.

2. 

Dräger Polytron Pulsar 2 Open Path Gas Detector is immune to sunlight so 
there is no need to take account of the sun position when siting detectors.

3. 

The density of the gas to be detected has to be considered. Methane is lighter 
than air and may be expected to rise, unless released at a low temperature 
or in a mixture with a heavier gas like carbon dioxide. Likewise, heavier 
hydrocarbons may be expected to fall. However such simple considerations 
as buoyancy may not be a reliable indicator of gas movement in practical 
situations. Gas leaking from high-pressure systems entrains with it a much 

larger volume of ambient air, forming a mixture that may be flammable and 

almost neutrally buoyant. In these circumstances, it is the natural air currents 
or forced ventilation that control the motion of the plume or cloud. Where 
air movements are unpredictable it may be necessary to use separate beam 

paths to cover different possibilities.

4. 

The distance between the transmitter and receiver should agree with the 
model selected (i.e. 4-60m, 30-120m or 100-200m). Note that the optical 

attenuator should be fitted below 16 metres.

5. 

The beam path and immediate surround should be kept free of obstructions 
that might hinder the free movement of air in the protected area or block the 
infrared beam. A clear path of 25cm diameter or greater is recommended. 
For maximum availability it is also recommended to avoid the following:

— Smoke stacks, chimneys and exhausts.
— Steam vents and plumes.

Summary of Contents for Polytron Pulsar 2

Page 1: ...Gas Detector en WARNING You must read understand and follow these instructions for use before you use the gas detector in order to ensure the proper operation and function of the gas detector Polytron...

Page 2: ......

Page 3: ...ding and strict observation of these instructions This gas detector is only to be used for the purposes specified herein Maintenance The device must be inspected and serviced by experts at regular int...

Page 4: ...e location Choosing the path of the beam and Mechanical Installation 1 The siting of an open path gas detector is often not as critical as a point detector since the released gas only has to find its...

Page 5: ...Polytron Pulsar 2 Open Path Gas Detector should be mounted to a stable structure free of excessive vibration Good choices would be a steel bulkhead brick wall concrete post or a rigid steel structure...

Page 6: ...5 5 6 6 7 7 S S E E 1 1 2 2 3 3 4 4 5 5 6 6 7 7 S S E E DRAWING NO GD8 59 TITLE Connection Diagram Ex e Terminal Box OPTION 2 powered via Receiver PRODUCT GD8 ISSUE NO 1 6 24 11 2011 Not for manufact...

Page 7: ...given in Figure 6 Place the socket components outer cap thrust washers outer seal etc onto the cable whilst ensuring that the cable armour braid is fitted between the clamp cone and sleeve Fit the con...

Page 8: ...o turn the socket insert it should not rotate If the insert does rotate then it is not fitted correctly and requires adjustment to secure it into position Finally secure the remaining components of th...

Page 9: ...display When aligned tighten the eight transmitter adjustment socket head screws and fully clamp Disconnect the Hand Held Terminal from the transmitter and reconnect to the receiver Figure 8 AP800 Att...

Page 10: ...enu then select ZERO A message is displayed SAFETY WARNING Are you sure beam path is free of gas If you are sure there is no gas present select YES otherwise press any other key to abort zero Once YES...

Page 11: ...n be written from the Hand Held Terminal with USER in Rx Memory Menu To see the settings that are available to be sent choose SETS in the Not Connected menu To change them connect the Dr ger Hand Held...

Page 12: ...for p n 2350325 The Hand Held Terminal The Dr ger Ltd Hand Held Terminal is a robust and weatherproof unit certified for use in hazardous areas It replaces the MTL611B Communicator based on a Psion O...

Page 13: ...ld Terminal or Communicator and has an infra red signal to a Data Wand This multiplicity of functions allows the AI500 to be used in systems with different levels of complexity In a fully engineered i...

Page 14: ...onal regulations WARNING The Dr ger Polytron Pulsar 2 Open Path Gas Detector contains no user serviceable components In the event of suspected failure of the transmitter or receiver the suspect unit s...

Page 15: ...ns with Safety Integrity Level SIL and any other configuration when applicable observe the Technical Manual How to Dispose of the Device From August 2005 regulations apply to the whole of the EU for t...

Page 16: ...0 C bis 60 C IECEx Certificate Number IEC Ex SIR 04 0006 Ex d ia IIC T5 Ta 40 C to 60 C Ex d ia IIC T6 Ta 40 C to 40 C UL Classi fied Class I Div 1 Groups C D CSA NRTL C Class I Div 1 Groups C D FM AN...

Page 17: ......

Page 18: ...lyth Northumberland NE24 4RG United Kingdom Tel 44 1670 352 891 Fax 44 1670 544 475 www draeger com Dr ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 ww...

Reviews: