Dräger Polytron Ex Instructions For Use Manual Download Page 8

 

Hints for Safe Use

 

Page 8 of 40

 

9023744  -  3rd edition - October 2005

 

Certain substances in the atmosphere to be monitored may impair the sensitivity of the sensing heads with Ex-sensor PR M resp. 
Ex-sensor LC M.
The following substances are known at present:
a) Polymerizing substances such as Acrylo nitrile, Butadiene and Styrene,
b) Catalyst poisons such as sulphurous or phosphorous compounds, halogenated hydrocarbons, silicon compounds, and metal-

organic vapours.

If such substances are expected to occur in the atmosphere to be monitored, the calibration intervals should be shortened. This 
is also recommended by the European Standard EN 50 073.

 

Intrinsic Safety Concept:

 

— Explosion protection acc. to CENELEC: 

— Sensor replacement and maintenance is possible without switching off the supply voltage.

The transmitters Polytron Ex resp. Polytron Ex R with sensing head SE Ex is marked by the device category 

 

II 2G

 

 and is suitable 

to be operated in hazardous zones 1 and 2.
The ex-approval is valid for the transmitters resp. sensing heads being operated in mixtures of flammable gases and vapours in 
air. Especially the ex-approval is not valid for the operation in oxygen enriched atmospheres.

 

Transmitter with measurement function for explosion protection according to 94/9/EC

 

Only the transmitter Polytron Ex (order no. 83 17 517) mit dem Ex-Sensor PR M (Sachnummer 68 09 225) with the Ex-sensor 
PR M (order no. 68 09 225) has been certified in the 1st supplement to the EC type test certificate BVS 03 ATEX E 160 X as 
suitable for use with the following gases and vapours as regards the measurement function for explosion protection:

Methane, propane, hydrogen, n-butane, ethene (ethylene) dimethylether, acetone, ammonia, gasoline 065/095 (FAM petrol), 
benzene, 1,3-butadiene, cyclopropane, diethylether, acetic acid-n-butylester (ethyl acetate), ethanol, ethyl acetate, ethylene oxide, 
n-hexane, methanol, methyl ethyl ketone (MEK), n-nonane, n-octane, 2-propanol (isopropanol), propene (propylene), toluene, 
acetylene, propylene oxide and n-pentane.

The following special conditions must be observed for safe use when using as a transmitter with a measurement function for 
explosion protection:
1. The heat-of-reaction measuring principle, based on catalytic oxidation of a flammable gas, is not absolutely reliable because 

at high measured gas concentrations the oxygen concentration in the sensor is no longer sufficient to oxidize the flammable 
gas. The measuring signal is thus diminished at very high gas concentrations and may reach values inside the measuring 
range. The downstream controller must be operated with display devices and measured value outputs (if implemented) and 
alarm outputs which lock if the measuring range is exceeded.

2. The heat-of-reaction measuring principle requires a minimum oxygen content of 12 % V/V. Otherwise too low measured values 

are displayed, due to the oxygen deficiency.

3. When the supply voltage is switched on, the transmitter may briefly (approx. 100 ms) send a current up to 12 mA to the central 

unit, exceeding the local alarm thresholds. To avoid this happening, the alarms at the central unit should be locked during the 
switch-on procedure.

Transmitter Polytron Ex:

EEx me [ib] IIC T4

Transmitter Polytron Ex R:

EEx me [ib] IIB T4

Summary of Contents for Polytron Ex

Page 1: ...9023744 3rd edition October 2005 Page 1 of 40 D Polytron Transmitter Ex and Ex R Instructions for Use 00123744_1 eps...

Page 2: ...between Transmitter and Central Unit 11 3 Wire Connection 11 4 Wire Connection 12 Connection between Transmitter Polytron Ex R and Sensing Head SE Ex 13 Installing Ex sensor in the Transmitter 13 Adju...

Page 3: ...on Ex and Ex R 24 Ex sensors 25 Ex sensor PR M 25 Ex sensor LC M 25 Sensing Heads SE Ex 26 Dimensions 26 Measuring other gases and vapours 27 Transmitter Ex Ex R 28 Sensing Head SE Ex 29 Design and Mo...

Page 4: ...cording to the national or European regulations on electrical equipment in potentially explosive atmospheres may be used only under the conditions speci ed here Modi cations of components or the use o...

Page 5: ...g range of 0 to 99 or 0 to 9 9 LEL can be realized Remark The measuring range 0 to 9 9 LEL is displayed without a decimal point as an integer value between 0 and 99 The sensing head SE Ex PR M ET resp...

Page 6: ...e point of detection are well accessible for the user Transmitter Polytron Ex with Ex sensor PR M Transmitter Polytron Ex with Ex sensor LC M Transmitter Polytron Ex R with remote sensing head SE Ex t...

Page 7: ...olytron Ex R The transmitter Polytron Ex R is intended to be used for the detection of ammable gases and vapours under atmospheric conditions The sensing head is connected to an inseparably connected...

Page 8: ...e for use with the following gases and vapours as regards the measurement function for explosion protection Methane propane hydrogen n butane ethene ethylene dimethylether acetone ammonia gasoline 065...

Page 9: ...cated above a possible leak or at the highest points at which major concentrations of gas may be found In case of gases and vapours with a density greater than that of air the sensing head resp transm...

Page 10: ...e pre conditioned connection cable bet ween sensing head SE Ex and transmitter Polytron Ex R Mount with four screws through the housing or by means of the mounting brackets Mounting the Sensing Head S...

Page 11: ...over and fix it with the screws Observe the tightness of the rubber gasket Complete installation by fixing the cover of the transmitter with four screws and observe tightness 3 Wire Connection The cab...

Page 12: ...itter terminals K1 and K3 must not exceed 20 Ohms The sum total of the input resistance of the central controller and the line resistances of both cores for the connection to the transmitter terminals...

Page 13: ...he ex approval Installing Ex sensor in the Transmitter 1 Loosen hexagon socket head bolts on upper side and remove upper section 2 Unscrew bayonet ring remove cover plate 3 Remove Ex sensor from packa...

Page 14: ...t points Switch on power supply 5 Using potentiometer adjust voltage at the test points for 270 mA 891 mV for 276 mA 911 mV 4 Disconnect voltmeter 6 Screw upper section onto housing using the hexagon...

Page 15: ...ses or vapours present or with calibration adapter Supply clean air at a rate of approx 0 5 L min via the calibration adapter 2 Wait for the display to stabilize max 3 minutes Set potentiometer such t...

Page 16: ...sensor must be replaced refer to page 22 Following Completion of Calibration 1 Set sliding switch to position 0 normal 2 Use special key to close flap in front of adjusters Calibrating Transmitter Po...

Page 17: ...ng in LEL For calibration with solvent vapours use can be made of the calibration chamber see Order List page 31 in which a specified amount of solvent is evaporated so as to obtain for example 50 LEL...

Page 18: ...mended calibration gas con centration s table below at a flow rate of approx 0 5 L min via the calibration adapter The calibration gas must be a mixture of the gas component to be monitored in air Mix...

Page 19: ...9023744 3rd edition October 2005 Page 19 of 40 Start Up Following Completion of Test 1 Set sliding switch to position 0 normal Use special key to close flap in front of adjusters 02823744_1 eps 1...

Page 20: ...List order no 68 10 537 This simply presses into place in the opening in front of the sensor sintered metal disc and locks in position The dust lter has virtually no adverse effect on sensor response...

Page 21: ...as warning system not exceeding a time interval of six months Check the zero adjustment and sensitivity calibration see page 15 to page 17 In particular it is important to check the condition of the s...

Page 22: ...cally safe The transmitter must be calibrated following sensor replacement see page 15 Allow 10 minutes for new Ex sensor to warm up Transmitter Polytron Ex R Use only Dr ger Sensors resp sensing head...

Page 23: ...ause Remedy No Display Cable faulty Check cable to central controller Transmitter resp sensing head can no longer be calibrated Ex sensor faulty or poisoned Replace Ex sensor page 22 Measurement curre...

Page 24: ...l conditions during operation 40 o C to 65 o C 2 700 to 1300 hPa in hazardous areas 800 to 1100 hPa 10 to 95 r h 2 The readability of the LC display might be affected at temperatures below 20 o C duri...

Page 25: ...y D 23560 L beck Germany Type Ex sensor PR M II 2G EEx d IIC T6 T5 T4 T3 DMT 97 ATEX E001X C0158 year of manufacturing by serial number 1 1 The year of manufacturing is coded by the third capital lett...

Page 26: ...r of manufacturing by serial number 1 1 The year of manufacturing is coded by the third capital letter of the serial number on the type plate S 2002 T 2003 U 2004 W 2005 X 2006 Y 2007 Z 2008 A 2009 B...

Page 27: ...methylether CH3OCH3 3 0 4 2 Ethanol C2H5OH 3 3 4 0 99 51 55 Ethene ethylene C2H4 2 7 5 2 Ethyl acetate CH3COOC2H5 2 1 4 0 99 67 72 Ethylene oxide C2H4O 2 6 3 0 51 36 41 2 n hexane C6H14 1 2 4 0 99 67...

Page 28: ...Technical Data Page 28 of 40 9023744 3rd edition October 2005 Transmitter Ex Ex R 106 mm 82 mm 152 mm 03823744_1 eps...

Page 29: ...9023744 3rd edition October 2005 Page 29 of 40 Technical Data Sensing Head SE Ex 45 mm 68 mm 03923744_1 eps...

Page 30: ...t pellistor The oxygen required for combustion is obtained from the monitored air The detector element is additionally heated by the resultant heat of reaction This heat in turn cau ses a change in th...

Page 31: ...ge 0 to 99 LEL 68 09 758 Sensing head SE Ex LC M measuring range 0 to 9 9 LEL 68 10 486 Accessories Cable gland 7 to 12 mm for second connecting cable 83 16 563 Dust lter 2 packing qty 10 2 Not tested...

Page 32: ...EC Type Examination Certi cate BVS 03 ATEX E 160 X Page 32 of 40 9023744 3rd edition October 2005 EC Type Examination Certi cate BVS 03 ATEX E 160 X...

Page 33: ...9023744 3rd edition October 2005 Page 33 of 40 EC Type Examination Certi cate BVS 03 ATEX E 160 X...

Page 34: ...Page 34 of 40 9023744 3rd edition October 2005...

Page 35: ...9023744 3rd edition October 2005 Page 35 of 40 EC Type Examination Certi cate BVS 03 ATEX E 160 X...

Page 36: ...Declaration of Conformity Page 36 of 40 9023744 3rd edition October 2005 Declaration of Conformity...

Page 37: ...alibration gas concentration 15 17 18 Catalyst poisons 8 CE marking 24 Central controller 5 Completion of Calibration 16 17 Completion of Test 19 Connection between Transmitter and Central Unit 11 Con...

Page 38: ...ts 9 10 One man calibration 5 Operating voltage 24 Operational parameters 25 26 Operational Use 20 Order List 31 Power consumption 24 Power supply circuit 25 Principle of Operation 30 Resolution 24 Sa...

Page 39: ...lytron Ex R with sensing head SE Ex LC M 6 Transmitter Polytron Ex R with sensing head SE Ex PR M ET 6 Transmitter with measurement function for explosion protection according to 94 9 EC 8 Ventilation...

Reviews: