34
Dräger Polytron 8100
Maintenance
6.3.2
Warning reference
Warning
number
Cause
Remedy
#101
Data-logger in Stack mode is 100 % full and
is not logging any more data.
Download the data, clear the data-logger.
#102
Data-logger in Stack mode is 90 % full.
Download the data as soon as possible and clear the data-
logger.
#103
#106
Data error in the instrument. Some functions
like data-logger, sensor test, etc. may no
longer be available.
Reset the instrument to factory default settings (see Section 7
on Page 36).
If this error occurs again:
Have the instrument checked by DrägerService.
#104
Date or time setting not valid.
Set date and time (see Section 5.8.3 on Page 22).
#107
Battery for data memory empty.
Have the instrument checked by DrägerService.
#110
#111
#112
SW dongle has been removed without being
deactivated.
Deactivate SW dongle (see Section 5.11 on Page 29).
#113
Alarms inhibited.
Enable alarms (see Section 5.8.1 on Page 21).
#164
#165
Sensor warm-up phase has not ended.
Increased measurement error must be
expected.
Wait until the sensor has warmed up. Do not calibrate before
sensor is fully warmed-up.
#167
#170
Calibration interval expired.
Recalibrate the instrument (see Section 6.2 on Page 30).
#171
Negative reading. Value below range
minimum.
Recalibrate zero point.
Summary of Contents for Polytron 8100
Page 2: ......
Page 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Page 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Page 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Page 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Page 253: ......