28
Dräger Polytron 5100
Technical data
9.5
Physical specifications
Enclosure
material
copper free aluminum or 316 stainless steel
Enclosure
protection
NEMA 4X and IP 65/66/67
Display
4 digit, 9 segment, back-lit
Size (LxWxD,approx.)
Without Docking Station
11.3” x 5.8” x 5.1”
(285 x 150 x 130 mm)
With Docking Station
11.5” x 7.1” x 7.4”
(295 x 180 x 190 mm)
Weight (approx.)
Without Docking Station
aluminum
4.9 lbs. (2.2 kg)
stainless steel 316
8.8 lbs. (3.6 kg)
With Docking Station
aluminum
7.7 lbs. (3.5 kg)
stainless steel 316
11.9 lbs (5.4 kg)
9.6
Environmental parameters
Pressure
20.7 to 38.4 in. of Hg (700 to 1300 hPa)
Humidity
0% to 100% RH, non-condensing
Temperature -40 to +149 °F (-40 to +65 °C)
Storage
temperature
-4 to +149 °F (-20 to +65 °C)
See sensor data sheet for sensor specifications
9.7
Environmental influences
For influences on the measurement performance and
restrictions of a particular sensor see sensor data sheet.
9.8
Tightening torque for instrument threads
The values are valid for aluminum and stainless steel 316
versions.
Part
TQ Lb. In.
TQ Nm
Enclosure lid
min. 44.3
min. 5
Sensor
min. 266
min. 30
Blind plug
min. 266
min. 30
Feed-through
Conduit hub
min. 443
min. 50
Instrument to
Docking Station
(4 screws)
71
8
Grounding screws
10.6
1.2
9.9
Tightening torque and size for field
wiring
terminals
Terminal
TQ Lb. In.
Wire Size
AWG
Wire Size
mm
2
Power
supply and
signal
4.4 – 7.0
(0.5 – 0.8 Nm)
24 – 12
0.2 – 2.5
Relays
4.4 – 7.0
(0.5 – 0.8 Nm)
20 – 12
0.5 – 2.5
www.norrscope.com