Attachment of carrying strap on back
event of constant mobile use outside the
of device
rescue vehicle the carrying strap can
If the carrying strap is constantly in
be secured in position in the slots in the
e. g. including cases where Oxylog is
Oxylog housing. See attachment
accommodated in holder and in the
gram in Fig. 7.
An inlet pressure at the device
The functioning of the Oxylog can be
pressed-gas connection 13 in Fig. 2) of
checked with the compressed-gas
at least 2 bar at a flow rate of 60
is
source (see
o n
required for operation of the Oxylog.
page 11).
The drive sources used (central gas
supply or gas cylinder with pressure
reducer) must always comply with the
aboveprerequisite.
Operation from a central gas supply
Screw
connecting hose (3 m or 5
Any upstream closing or metering
m, see Order List) to device and insert
elements must be
fully open!
connector into wall outlet valve.
8