Dräger Oxylog Instructions For Use Manual Download Page 13

Use in toxic atmosphere

The Oxylog can also be used for

controlled 

ventilation of injured per-

sons in a toxic ambient atmosphere. The
switch setting 

 Air 

 is to be

employed for this purpose.

If the patient breathes spontaneously or
if 

spontaneous breathing 

has been

restored after resuscitation the partially

spontaneous intake of toxic ambient air

is 

not 

prevented by the Oxylog.

It is for this reason that the special
Oxylog with spontaneous breathing
device (cf. Order List) must be used.
This device facilitates spontaneous
breathing with an airway pressure 0 to
be carried out in addition to controlled
ventilation (pay attention to respective
Operating Manual).

Note
In the case of spontaneous breathing

the mask must always fit snugly, so as to

prevent the intake of toxic ambient air.

Ventilation with PEEP

(Special accessory)

Pay attention to respective Operating
Manual.

Set PEEP valve to 0 and push it onto the
waste-gas socket of the 
ing valve:
Determine inspiratory airway pressure
at airway pressure gauge.

Set PEEP value:

The endinspiratory pressure increases
approximately by the set PEEP value.

Should the airway pressure rise higher
or not change its value at all, the PEEP
valve is defective and must be replaced.

Upon termination of the PEEP mode the

PEEP valve must be removed from the

valve of the airway pressure gauge.

Use of pressure limiting valve

(special accessory)

Set rotary knob at optional pressure
limiting valve (located on inspiration
socket at Oxylog) and check inspiration

pressure limited in this manner on
airway pressure gauge.

Caution!
In the case of pressure-limited ventila-
tion the set MV does not have the full
effect. It is advisable to set the pressure

limitation roughly 10 mbar in excess of

the inspiration pressure, so that the
pressure limitation only becomes effec-
tive in exceptional circumstances (e. g.

coughing).

Use of expiratory volume measure-
ment 

(special accessory)

Insert hose nozzle into waste-gas
socket of non-rebreathing valve.
Connect Volumeter 3000 to silicone
tubing (1.1 m).

The Volumeter 3000 can be used to

measure both the tidal volume and the
MV (pay attention to the respective
Operating Manual).

Caution!
Expiratory volume measurement is not

possible when using the PEEP valve.

 

Set main switch to 

 When using

compressed-gas cylinder, close cylin-
der valve.

Summary of Contents for Oxylog

Page 1: ...Oxylog Emergency Ventilator Instructions for Use Deutscher Text Bitte umdrehen...

Page 2: ...not conforming to its untended use Dragerwerk Akt engese schaft cannot be he d responsible for damage caused by non compliance with the recommendations given above The warranty and ab hty provisions...

Page 3: ...rdiopulmon ary resuscitation ratio 12 min 4 Rotary knob for setting minute volume MV 5 Pneumatic main switch I O 6 Switch asAir Mix No Air Mix for transferring patients by road or air for ventilation...

Page 4: ...nlets and outlets 11 Socket for ventilation tubing 13 Compressed gas connection male 12 Filter for purifying ambient air thread M 15 x 1 drawn in 14 Slot for alternative fastening of carrying strap e...

Page 5: ...h patient connec tion and waste gas socket The patient connection features an external taper ISO dia 22 and an internal taper ISO dia 15 T h e waste gas socket has an internal taper ISO dia 22 for con...

Page 6: ...evice into holder such that stems of two studs 7 on housing slide into slots Fig 4 Holder 6404560 without bracket 6 16a of holder Studs must engage in hole in brackets 17 Press on brackets 17 to ensur...

Page 7: ...Fig 5 Oxylog in holder Fig 6 Holderwith rail bracket...

Page 8: ...tion An inlet pressure at the device com The functioning of the Oxylog can be pressed gas connection 13 in Fig 2 of checked with the compressed gas at least 2 bar at a flow rate of 60 L min is source...

Page 9: ...iation flights low defined O2 concentrations may be required For this purpose a compressed gas blender can be connected upstream of the Oxylog see Order List Caution The inlet pressure at the Oxylog m...

Page 10: ...m Silicone tubing E 1 l m l Non rebreathing valve Ambu PEEP valve Only for operation with extendedventilationhose 1Fig 9 Oxylog with patient system 2 Fig 10 Individual components of non rebreathing va...

Page 11: ...L min Airway pressure 28 to 38 mbar The Oxylog should switch at regular intervals from inspiration to expiration The operating prerequisites at least 2 Functional Check bar at 60 L min at Oxylog are...

Page 12: ...to 80 vol Connect non rebreathing valve to mask or tube Make sure that airways are completely free Check airway pressure gauge In normal circumstances the inspiratory airway pressure values should be...

Page 13: ...sure increases approximately by the set PEEP value Should the airway pressure rise higher or not change its value at all the PEEP valve is defective and must be replaced Upon termination of the PEEP m...

Page 14: ...ched in parallel for supply of spontaneous breathing Without Air Mix switch Non rebreathing valve with expiratory check valve The following configuration is a prerequisite for use in a toxic atmospher...

Page 15: ...contact with the edge of the housing Fit the cover by applying slight pressure and turn clockwise by about 45 Make sure that the diaphragm sits smoothly in the housing Pull the perforated cap at the o...

Page 16: ...CPAP is not possible Care Stripping down 0 Detach ventilation hose from Oxylog 0 Detach non rebreathing valve from hose 0 Detach mask from non rebreathing valve 0 Turn cover of non rebreathing valve...

Page 17: ...al dia 6 mm 100 mm lona 3 Cap of check valve 7 Disposable syringe IO mL Detach cap from check valve Attach syringe together with catheter connector Seal inspiratory port with sealing plug Using the sy...

Page 18: ...g valve with 5 Pressure gas line 5 bar check valve 84 08 568 6 Cap nut 84 08 298 2 Ventilation hose 84 04 063 7 Screw plug 84 08 299 3 Demand valve 8 Mouthpiece 84 07 303 4 Oxylog Open oxygen cylinder...

Page 19: ...l section Carefully remove residual water from Volumeter All parts are to be thoroughly dried following cleaning and rinsing The component parts can also be washed in the Drager Purfactor which automa...

Page 20: ...g can be sterilized at 134 C The Oxylog cannot be sterilized Plastic and rubber mouldings are to be disassembled prior to sterilization Following sterilization in an autoclave the device is to be disa...

Page 21: ...g phragm in non rebreath valve and assemble ing valve deformed or correctly out of shape Device comes to a halt on inspirati0r r Inadequate supply pressure Oxylog defective Establish adequate supply p...

Page 22: ...100v0l 02 50 mbar to 80 mbar Pressure gauge 1Oto 80mbar 2 5 full scale Drive gas Quality Pressure at device inlet O2 or air Dry oil free and dust free from a central supply unit or from com pressed ga...

Page 23: ...relative humidity 600 l 200 m bar Materials used Oxylog housing Impact resistant ABS Acrylonitrile Butadiene Styrene Ventilation tubing Silicone rubber Non rebreathing valve Housing PSU Polysulfone Co...

Page 24: ...or B 03 580 O2 cylinder AG 1 l 200 G 314 straight valve filled and B 02710 OS compressed air connecting hose 1 5 m or M 17816 O2 compressed air connecting hose 3 m For operation from central supply u...

Page 25: ...air connecting hose 5 m angled plug connector M 22 495 Special accessories Silicone mask No 2 Silicone mask No 5 Holder to secure Oxylog in mobile units Holder with rail bracket to attach Oxylog to wa...

Page 26: ...size 5 stainless steel For operation with extended ventilation hose Connection hose silicone 1 1 m Socket IS0 Spare and wearing parts for sterilisation Non rebreathing valve 2 Connection hose Silicone...

Page 27: ...52 5 Valve housing 84 06 595 67 Carrying strap 84 04 773 6 Strap 84 04 078 7 Buckle 8405 179 8 Pressure limiting valve 84 05 390 9 PEEP valve Ambu O 10 mbar 84 07 475 10 Silicone mask No 2 21 20 194 1...

Page 28: ...17 a 13 izv I F I L Fig 1 1 Gomponent parts see rans LISIJ Optional only for operation with extended ventilation hose The item Nos are not identical with the item Nos in the other figures 28...

Page 29: ...for Use are provided for general information only and are not intended for use with any specific machine or device Drlgerwerk Aktiengesellschaft Germany Moislinger Allee 53 55 D 23542 Liibeck 3 4 5 1...

Reviews: