3.  Installation and commissioning

The device was thoroughly cleaned before leaving the factory. However, due to the transport con-

ditions and the transport packaging, neither a sterile nor a disinfected product can be assumed 

upon delivery. For the initial cleaning, please observe the instructions provided in 

Section 5.1

.

First check the device and accessories for completeness by comparing with the enclosed delivery 

bill.

The device may only be operated in an upright position; operation with an inclination of more than 

30° from the normal in one direction is not permitted. Operation in inadmissible positions (upside 

down on its head, lying on its back or on its display) is not permitted and may result in damage to 

the device.

Only use the screws supplied. Always use all fastening points for safe mounting of the device.

Adequate ventilation is necessary for reliable and continuous operation of the device. Ventilation 

protects the device from overheating. Therefore, do not install the device in a location or under 

conditions where adequate ventilation of the housing surfaces is not possible. To avoid heat accu-

mulation or interruption of air circulation and to protect the device mounting from possible over-

loads, it is not permitted to place objects or materials on the device or to hang them on the device.

To avoid the risk of injury and damage to the device, all connection cables should be routed so that 

no one can step on them, trip over them, or place heavy objects on them. When mounting, ensure 

a fixed position and/or stable suspension.

3.1 Cable and device connection

Use only the supplied connection cables.

First connect the potential equalization cable (special green-yellow marking) to ensure sufficient 

potential equalization between user and device. Operation in the medical field without connected 

potential equalization is not permitted.

Before connecting to a supply network, check the correct voltage and frequency of your power 

supply network (see nameplate on the housing). Carefully connect all desired connection cables 

(power cable, keyboard, mouse, serial cable, network, etc.) to the device. To avoid the risk of elec-

tric shock, this device must only be connected to a supply network with a protective earth conduc-

tor. Finally, check the tight fit of all cable connections. It is not necessary for the end customer to 

attach strain reliefs to the device.

For complete disconnection from the mains, the system mains switch must be actuated or the 

mains plug must be pulled out of the socket. When disconnecting from the mains, always grasp 

the mains cable by the plug. Do not pull on the cable. Do not place the device in conditions that 

make it difficult to disconnect it from the mains power supply.

Summary of Contents for OR-PC 21

Page 1: ......

Page 2: ...Electromagnetic compatibility EMC 7 3 4 BIOS 9 3 5 Graphic symbols 9 4 Operation 10 4 1 Keypad 10 5 Maintenance 11 5 1 Care and hygiene instructions 11 5 2 Maintenance intervals and wearing parts 12 5 3 Malfunctions and alarm signals 12 5 4 In house repairs 13 5 5 Transport notes for ACL devices 14 5 6 Environmentally sound disposal 14 6 Manufacturer contact 15 7 Warranty and guarantee 15 8 Applic...

Page 3: ...ched environment is not intended 6 6 The device is suitable for continuous operation When connecting devices it is not permitted to touch the patient and the contacts of connectors of the signal input or output parts at the same time The device is a professional ME device as defined in IEC 60601 1 2 2014 para 3 23 suitable for use within the patient environment not suitable for use with direct con...

Page 4: ...place objects or materials on the device or to hang them on the device To avoid the risk of injury and damage to the device all connection cables should be routed so that no one can step on them trip over them or place heavy objects on them When mounting ensure a fixed position and or stable suspension 3 1 Cable and device connection Use only the supplied connection cables First connect the potent...

Page 5: ...these decoupled interfaces At the same time a functional unit is created when exchanging data with other devices This applies in particular if this functional unit has a medical purpose e g control of infusion pumps by the device connection of application parts The validity of any assured standard conformities and properties of the device would not extend to the new functional unit Instead the per...

Page 6: ... to 19 kPa Store the device for three hours at its new operating location before commissioning if the tem perature difference during transport is more than 10 C Rapid temperature changes can cause damage to the device e g due to condensation Avoid displaying the same screen content for an extended period of time Otherwise TFT displays may show afterimages at least temporarily image sticking effect...

Page 7: ...s other than residential and those directly connected to the public utility network which also supplies buildings used for residential purposes Voltage fluctuation flicker emissions according to IEC 61000 3 3 complies Nominal power of the transmitter in W Protective distance depending on the transmitting frequency in m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 7 GHz 0 01 0 12 0 12 0 23 0 1 ...

Page 8: ...ut monopolar coagu lation spray 3 3 1 FCC Statement This unit has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This unit generates uses and can radiate radio frequency energy and if not installed and used in accor...

Page 9: ...NSI AAMI ES60601 1 A1 2012 C1 2009 I2012 und A2 2010 I2012 CSA CAN CSA C22 2 NO 60601 1 14 Underwriter Laboratories LLC EAC mark CU TR 004 2011 This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules FCC ID 2AUWB2 47 CFR Part 15 Industrial protection IEC 60529 Observe manual instructions ISO 7010 M002 Do not open device n...

Page 10: ...aries depending on the selected equipment variant The following key functions are integrated Function keys LED indicators OEM Dräger Standby Taster LED seitlich neben Taste Stand 15 12 2020 OEM Dräger Standby Taster LED innerhalb des Schlosses Stand 15 12 2020 Switch on off Shutdown in standby OEM Dräger Standby Taster LED seitlich neben Taste Stand 15 12 2020 OEM Dräger Standby Taster LED innerha...

Page 11: ...re time in minutes Example products Alcohol 100 360 Bacillol 30 Foam Meliseptol Biguanide 8 360 Incidin Plus Chlorine organic or inorganic substances with active chlorine 10 360 Incidin perfekt Optisept Alkaline solution 8 360 Milk of lime Per compounds 8 360 Incidin active Terralin PAA Phenol or phenol derivatives 6 360 Helipur The manufacturer s instructions for the disinfectants must be observe...

Page 12: ...nals If any mechanical damage to the housing display and connection cables is detected the device must be taken out of operation In this case the device must be checked by service personnel Operation with damaged components can lead to further defects as well as data loss In the following cases restart the device and check the device environment for strong sources of EMC interference and for non c...

Page 13: ...r training service manual and repair assistance please contact the manufacturer contact listed in Section 6 Improper repairs pose a safety risk The use of your own setting or adjustment methods that do not comply with the device specifications or of components not supplied or approved by ACL is not permitted and may result in damage to the device Any changes to the device without the manufacturer ...

Page 14: ...ust not be disposed of together with unsorted munici pal waste household waste Separate collection is a prerequisite for reuse recycling and recovery of waste electrical and electronic equipment thus conserving resources For equipment return please contact the manufacturer s address given in Section 6 or an appro priately authorized disposal company We will be happy to advise you on the options fo...

Page 15: ...dealer occurred The Convention on Contracts for the International Sale of Goods does not apply 9 Data protection We would like to point out that data collected from you in the course of processing your order or preparing an offer will be electronically recorded processed and evaluated for the purpose of qua lity assurance We assure you that this data will be treated in accordance with the German G...

Page 16: ...ALWAYS A STEP AHEAD 17 EN ...

Reviews: