Measurement Mode
Procedures for measuring cardiac output differ according to the mode of measurement
you select. You ordinarily measure C.O. in automatic mode. If unstable blood
temperatures, artifact, or other conditions prevent an automatic measurement, you can
still take C.O. measurements by selecting manual mode. (
Manual
is the default
setting on the C.O. setup menu.)
In automatic mode, the
READY
message appears when the monitor determines that the
baseline blood temperature is stable. Do not make a C.O. injection before this message
appears (this message indicates that the monitor is ready to detect the injectate's
decreased temperature). An unstable blood temperature removes the READY
message. It does not reappear until the patient’s blood temperature is stable again.
In both manual and automatic modes, the monitor sounds an attention tone when the
C.O. value has been computed. On the C.O. Averaging screen, the value is displayed
in the next available box, and the
Save AVG
field is updated. The value in the main
screen parameter box does not change until you save the C.O. average.
To change the measurement mode, open the C.O. setup menu as described on page 15-
6 and select the desired mode. Procedures for measuring C.O. in automatic or manual
mode are described on the following page.
Injectate
Volume
Displays the currently selected volume of the
injectate used to measure cardiac output.
NOTE
: If
Other
is selected for
Catheter Type
,
this field is ghosted.
• 3.0
,
5.0
,
10.0 cc
Comp.
Constant
Compensates for discrepancies in catheters;
see page 15-8 for more detailed information.
Not applicable
Mode
Determines the mode of measurement for
cardiac output; see page 15-7 for more detailed
information
• Auto
• Manual
BT Alarm
Opens the Alarm Limits table beginning with
temperature parameters.
Not applicable
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...