![Dräger Infinity Delta Instructions For Use Manual Download Page 262](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779262.webp)
15-6
D
ELTA
/D
ELTA
XL/K
APPA
VF8
15 C
ARDIAC
O
UTPUT
(C.O.)
Cardiac output signals, injectate temperature (IT), and blood temperature (BT) can be
obtained from the 2-pressure pod (Hemo2) or the 4-pressure pod (Hemo4, or
MPod – Quad Hemo). You can acquire cardiac output signals but not body
temperature signals from the HemoMed.
C.O. Setup Menu
To access the C.O. setup menu
Click on the C.O. parameter box on the main screen.
or
1. Press the
Menu
fixed key to display the Main menu.
2. Click on
Patient Setup
.
3. Click on
Parameters
.
4. Scroll to
C.O.
and click to display the C.O. setup menu.
Quick Reference – C.O. Setup
NOTE:
Both IT and BT signals must be obtained from the same pod.
Menu Item
Description
Available Settings
C.O. Start
Starts a C.O. measurement (see page 15-10)
• Not applicable
Catheter Type
Displays the currently selected catheter type.
NOTE
: Due to corporate mergers, Baxter
cardiac output catheters and accessories may
be labelled as being from Edwards. Ohmeda
cardiac output catheters and accessories may
be labelled as being from Becton Dickinson
(BD). Contact Edwards and/or BD if there is any
doubt as to the identity of the cardiac output
catheters or accessories.
Click on one of the following to
change the catheter type:
• BD/Ohmeda
• Edw./Baxter
• Arrow
• Other
Catheter Size
Displays the currently selected catheter size.
NOTE
: If Other is selected for Catheter Type,
this field is ghosted.
Click on one of the following to
change the catheter size:
• 5, 7,
or
7.5 F
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...