Technical data
Device with power supply unit PS500
(8418950)
For charging subsequent to storage
For storage up to 6 months
–15 to 25 °C (5 to 77 °F)
For storage up to 3 months
–15 to 40 °C (5 to 104 °F)
Information on the batteries
The technical specifications of the battery man-
ufacturer regarding the maximum storage life
refer to a relative humidity of 45 to 85 %. If it is
stored outside of these conditions, a battery
test must be performed before the device is
used. Completely charging the battery every
6 months at the latest is recommended. Several
charge cycles and discharge cycles may be
required as a battery test in order to completely
reactivate the electrochemical composition after
long term storage.
Depending on the accessories used, more
stringent ambient conditions may apply.
Observe the corresponding instructions for
use.
8.2 Operating
characteristics
Batteries
The battery runtime applies to fully charged bat-
teries and during ventilation in accordance with
ISO 80601-2-12:2020 as well as at an ambient
temperature of approx. 22 °C (71.6 °F).
The battery runtime information after
50 discharge and charge cycles refers to the
shortest battery runtime under the test condi-
tions specified in the standard.
Low temperatures may reduce the battery run-
time.
The charging time applies to new and com-
pletely discharged batteries during ventilation
without GS500. The actual charging time
depends on the battery charge.
If the GS500 is operating or the ambient tem-
perature is high, the battery charging process
may be restricted or interrupted.
Internal battery of ventilation unit (without
PS500)
Type
NiMH battery, sealed
F3 fuse of the ventilation unit
F15A 80V UL248
8.1
Ambient conditions (continued)
Summary of Contents for Babylog VN Series
Page 21: ...Evita V Babylog VN 21 3 3 1 3 1 1 25 40813 1 2 3 4 2 3 1 4...
Page 22: ...22 Evita V Babylog VN 4 4 1 4 1 1 39417 1 2 3 1 2 3...
Page 24: ...24 Evita V Babylog VN 4 1 2 26 24 4 1 3 32 36 27 4 1 4 PS500 32...
Page 25: ...Evita V Babylog VN 25 5 5 1 5 1 1 27 28 5 1 2 PS500 PS500...
Page 26: ...26 Evita V Babylog VN 6 6 1 210 PS500 PS500 160 200 201 127 100 254 5...
Page 27: ...Evita V Babylog VN 27 7 7 1 120 100 250 10 251 10 3 28 NiMH PS500 VRLA 2 PS500 LFP 4...
Page 28: ...28 Evita V Babylog VN 7 2 7 2 1 32 32 5 5 7 2 2 GS500...
Page 33: ...Evita V Babylog VN 33 45 85 6 8 2 ISO 80601 2 12 2020 22 C 71 6 F 50 GS500 GS500 8 1 continued...
Page 36: ...36 Evita V Babylog VN NiMH VRLA LFP 12 Ah 2 1000 Ah 2 5 A GS500 9 1 3 25 49 0 24...
Page 39: ...Evita V Babylog VN 39 10 5 GS500 GS500 GS500...
Page 178: ...178 Evita V Babylog VN 3 3 1 3 1 1 182 40813 1 2 3 4 2 3 1 4...
Page 179: ...Evita V Babylog VN 179 4 4 1 4 1 1 39417 1 2 3 99 100 75 98 50 74 1 2 3...
Page 180: ...180 Evita V Babylog VN PS500 PS500 192 5 10 25 49 0 24 40757 1 2 PS500 3 4 3 1 2 4 3 4...
Page 181: ...Evita V Babylog VN 181 4 1 2 183 181 4 1 3 Te 189 193 184 4 1 4 PS500 Te 189...
Page 182: ...182 Evita V Babylog VN yp 5 yp 5 1 5 1 1 184 185 5 1 2 PS500 PS500...
Page 183: ...Evita V Babylog VN 183 6 6 1 C 160 200 201 210 PS500 PS500 254 5 100 127 250 10 251 10...
Page 184: ...184 Evita V Babylog VN 7 7 1 100 120 3 185 NiMH PS500 VRLA 2 PS500 LFP 4 C...
Page 185: ...Evita V Babylog VN 185 7 2 7 2 1 Te 189 Te 189 5 5 7 2 2...
Page 193: ...Evita V Babylog VN 193 NiMH VRLA LFP 12 2 1000 2 5 A GS500 9 1 3 25 49 0 24...
Page 195: ...Evita V Babylog VN 195 9 1 4 183 PS500 PS500 PS500 PS500 GS500 41052 36 36 6 12 18 24 1 2 2 1...
Page 196: ...196 Evita V Babylog VN 10 5 GS500 GS500 GS500...