27
Measurement accuracy
Reproducibility with an ethanol standard
up to 50 μg/100 mL
0.8 μg/100 mL or
≥
50 μg/100 mL
1.7 % of measurement value
the higher value is valid
up to 1.00 ‰
0.017 ‰ or
≥
1.00 ‰
1.7 % of measurement value
the higher value is valid
Sensitivity drift
typically 0.6 % of meas. value/month
Intervals before readiness for operation
1st measurement
about 6 s
after terminating a result of a previous test in range:
(breath alcohol unit)
25 μg/100 mL
1
20 s
50 μg/100 mL
1
40 s
100 μg/100 mL
1
80 s
> 150
μg/100 mL
120 s
Dimensions
140 mm x 80 mm x 35 mm (W x H x D)
Weight
approx. 195 g
Electrical supply
Batteries
2 x 1.5 V Type Alkaline
(Mignon, LR6, AA)
approx. 1500 measurements
CE-marking
electromagnetic compatibility
(Directive 89/336/EEC)