Measuring limits
Factory settings
Error
In case of internal error, the following error message is displayed.
If the failure still occurs after the instrument is switched on and off several times. Contact
Dräger Safety.
Limit
Displaymode
Set mode
μg/100 mL
<4
μg/100 mL
≤
9
μg/100 mL
<31
μg/100 mL
<36
μg/100 mL
≥
36
AIR
ZERO
PASS
FAIL
FAIL
FAIL
ROAD
ZERO
PASS
PASS
ALERT
FAIL
MARINE
ZERO
PASS
PASS
ALERT
FAIL
RAIL
ZERO
PASS
PASS
ALERT
FAIL
alcohol indication
ROAD
AIR
MARINE
RAIL
ZERO: <4 μg/100 mL
ZERO: <4 μg/100 mL
ZERO: <4 μg/100 mL
ZERO: <4 μg/100 mL
PASS: <31 μg/100 mL
PASS:
≤
9 μg/100 mL
PASS: <31 μg/100 mL
PASS: <31 μg/100 mL
ALERT:
≤
35 μg/100 mL FAIL: >9 μg/100 mL
ALERT:
≤
35 μg/100 mL ALERT:
≤
35 μg/100 mL
FAIL:
≥
36 μg/100 mL
FAIL:
≥
36 μg/100 mL
FAIL:
≥
36 μg/100 mL