22
Violation Reset
Too many violations have been detected
Call an authorized Dräger Interlock Service Center in
due time.
Handset error / Contact Service !
Fault of handset
Switch ignition off/on and repeat the test.
If message reappears: call an authorized Dräger Inter-
lock Service Center.
Controlbox error / Contact Service !
Fault of control box
Switch ignition off/on and repeat the test.
If message reappears: call an authorized Dräger Inter-
lock Service Center.
Check cable and reconnect !
No connection between handset and con-
trol box
Check plugs of the connecting cable and connect
again with handset and control box.
If message reappears: call an authorized Dräger Inter-
lock Service Center.
(No display textes on handset, but signal
lamps functioning)
Temperature of the handset too low
Perform test as usual guided by the signal lamps.
Detach handset and take it at a place with moderate
temperature. Connect handset again and make meas-
urement.
(No display messages on handset)
No connection between handset and con-
trol box
Check plugs of the connecting cable, disconnect plug
from handset and reconnect again.
If again no messages are shown: call an authorized
Dräger Interlock Service Center.
Display
Cause
Remedy