12
3.5
Output and Display Variations
The following table shows the status of the 4 to 20 mA output, and the LCD
display for various conditions.
Condition
Analog Output
LCD Display
Warming-up
actual signal
actual signal
After warm-up
actual signal
actual signal
After power outage
actual signal
actual signal
New sensor installed
actual signal
actual signal
Sensor removed
signal
“pls con snr”
Hardware fault
signal
“flt”
Microprocessor lock-up
signal
frozen display
In menu
maintenance signal
menu function
In cal modes
maintenance signal
cal function displays
The “actual signal” means the unmodified real-time signal in the circuit.
The maintenance signal is the 3 to 5mA oscillating signal, or the steady 3 mA
signal chosen in the “cal sig” menu function.