Safety-related information
Dräger Oxygen Booster 200 ECO
17
Other hazards
DANGER
Danger of death and risk of equipment damage due to
overfilling compressed air containers.
Compressed air containers warm up when being filled
and therefore have a higher pressure when warm.
When subsequently cooled, the pressure in the
compressed gas container drops, for which reason the
limit pressure switch is set to 215+3 bar/3150+43.5 psi
(type A) or 228+3 bar/3300+43.5 psi (type B)
depending on type.
After cooling, compressed gas containers are only
permitted to be filled up to 200 bar/2900 psi (type A) or
207 bar/3000 psi (type B).
DANGER
Danger of death and risk of equipment damage when
operating the Dräger Oxygen Booster 200 ECO
unsupervised.
The Dräger Oxygen Booster 200 ECO may only be
operated under supervision.
DANGER
Danger of death and risk of equipment damage due to
fire caused by oxygen enrichment.
Only connect the filling panels and hoses specified in
chapter "Order list" on page 25.
Do not use any adapters.
WARNING
Risk of injury and of equipment damage due to
incorrectly dimensionsed accessories.
The Dräger Oxygen Booster 200 ECO is only
permitted for use together with compressed gas tanks
approved for a high pressure range up to 207 bar/
3000 psi.
The accessories used for oxygen must also be suitable
and clean.
WARNING
Danger of death and risk of equipment damage due to
electric shock.
The switch box of the Dräger Oxygen Booster
200 ECO is not permitted to be damp.
Disconnect the Dräger Oxygen Booster 200 ECO from
the power supply immediately if a short circuit is
suspected.
!
!
!
!
!
WARNING
Risk of injury and of equipment damage when
operating the Dräger Oxygen Booster 200 ECO
outside the case housing.
The Dräger Oxygen Booster 200 ECO is only
permitted to be removed from the case housing for
servicing and maintenance purposes by authorized
specialist personnel; during all other work, the Dräger
Oxygen Booster 200 ECO must be securely fitted in
the case housing. Operation outside the case housing
is forbidden and there is a risk of injury due to moving
parts.
Check all screw connections and tighten if necessary
before every use of the Dräger Oxygen Booster
200 ECO.
WARNING
Risk of injury due to contaminated lines.
Use only pure oxygen or pure oxygen mixtures for
filling and compressing. Otherwise the piping system
may become contaminated.
A contaminated Dräger Oxygen Booster 200 ECO
must be cleaned before use by authorized expert
personnel or DrägerService.
Observe ASTM G93 class C.
CAUTION
Risk of injury and risk of equipment damage due to
high temperatures developing in the Dräger Oxygen
Booster 200 ECO.
In the event of high temperatures developing in the
Dräger Oxygen Booster 200 ECO, the pump may
overheat and limit the function of the Dräger Oxygen
Booster 200 ECO and could catch fire in some
circumstances.
Do not seal or block the ventilation slots in the Dräger
Oxygen Booster 200 ECO. A minimum distance of
30 cm to the surrounding walls must be observed to
avoid impaired cooling.
CAUTION
Risk of injury and risk of equipment damage due to
vibration of the Dräger Oxygen Booster 200 ECO.
The Dräger Oxygen Booster 200 ECO may vibrate
strongly in use and could fall down. Only use the
Dräger Oxygen Booster 200 ECO in a horizontal
position.
Always secure during use or use an antislip base.
CAUTION
Risk of injury and risk of equipment damage due to
high dead weight.
The Dräger Oxygen Booster 200 ECO weighs approx.
75 kg.
Always take care when transporting the Dräger
Oxygen Booster 200 ECO in order to avoid injury.
Always secure the Dräger Oxygen Booster 200 ECO
adequately during transport.
Securing the transport case for the filling panel for six
compressed gas cylinders, for example with cable ties
or a lock, is recommended.
CAUTION
Risk of tripping over hoses and cables.
Secure and mark horizontal or vertical hoses and
cables.
!
!
!
!
!
!