F
o
r int
e
rnal us
e only
.
Copyright
res
e
rved.
R
566
4900T
03.f
m
03.
12.01
Dräger Medizintechnik GmbH
Repair Information/Error Events Savina Version 2.0
Page 273
5HSDLU,QIRUPDWLRQ(UURU(YHQWV
.QRZQHUURUV
Check fuse for internal battery.
Replace fuse.
Check voltage of internal battery in Service
Mode
Is fuse OK?
Voltage within
tolerance?
Yes
No
Yes
No
Power pack faulty.
Voltage less
than 10 V?
Yes
No
Replace battery.
Remove internal battery (see Maintenance Procedure, chapter "
internal rechargeable batteries
"), recharge with external charger and re-install.
Or connect external battery (see
"Starting aid" for charging of internal battery
Observe charging specifications of battery.