Error list Savina Version 3.0
Page 122
(UURUOLVW
520IURQWSURFHVVRU)3HUURUJURXS
11.0211
Data retention: EEPROM write access
error
−
11.0212
Data retention: RTC-RAM write access
error
−
11.0300
Data retention (EEPROM), checksum
error: Operating minutes storage.
−
Switch on Savina again.
−
Replace EEPROM FP.
11.0301
Data retention (EEPROM), checksum
error: Data structure in EEPROM
(Magic).
−
Complete new data input in EEPROM
MP (see error codes 11.0300 and
11.0302 through 11.0306).
−
Replace EEPROM FP.
11.0302
Data retention (EEPROM), checksum
error: Software options (not used for
the time being).
−
11.0303
Data retention (EEPROM), checksum
error: DeviceID (not used for the time
being).
−
11.0304
Data retention (EEPROM), checksum
error: Settings:
−
Resetting of language version under
"Config".
−
Replace EEPROM FP.
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ